- 20 -
7.3.3 SpO
2
& Pulse Rate Measurement Value Verifi cation
a) Connect the SpO
2
Probe to the SpO
2
connector on the pulse oximeter.
b) Insert the operator’s
fi
nger into the
fi
nger probe, the SpO
2
measured value of a healthy
person should be from 95% to 99%, and the pulse rate is same as heart rate.
c) If the SpO
2
Simulator is available, verify the accuracy of the Oxygen Saturation Value
with probes as follows:
Oxygen Saturation Tolerance
96% ±2%
86% ±2%
70% ±3%
7.3.4 SpO
2
& Pulse Rate Alarm Verifi cation
a). Connect the SpO
2
Probe to the SpO
2
connector on the pulse oximeter.
b). Insert the operator’s
fi
nger into the
fi
nger probe, the SpO
2
measured value of a healthy
person should be more than 96%.
c). Set the SpO
2
high limit as 90, low limit as 80.
d). Verify the SpO
2
visual and auditory alarms. The color of the SpO
2
data should be red and
"dudu" voice should be heard.
7.4 Trouble Shooting
a) Can’t power on the pulse oximeter
Please check the batteries voltage.
b) "Probe OFF " alarm
Please check if the probe was connected with the pulse oximeter correctly. If the probe is
with extension cable please check if the extension cable is connected with the probe correctly.
7.5 Warranty and Repair
7.5.1 Maintenance Method
a)
Maintenance responding time:
9:00am ~ 17:30pm, Monday to Friday
b)
Service support:
Our company will offer user telephone and e-mail technology support
and parts change.
Parts change:
our company will change parts if it is necessary free of charge in the
warranty period.
Because parts are the sources of maintenance, user should send them back to our
company if not speci
fi
ed.
c) Update the system software free of charge.
7.5.2 Exemption and Limitation:
a) Our company isn’t responsible for such damage caused by force majeure. For example:
fi
re, thunder
fl
ash,
fl
ood, cyclone, hail, earthquake, house collapse, commotion, plane
failing and traf
fi
c accident, deliberate damage, lack of fuel or water, labor and capital
bother, strike and stop-working etc.
PULSE OXIMETER