AUDIBLE ALARM INHIBITION/OFF:
Short pressing the button to silence the audible alarm for 2 minutes, the audible alarm
indicator will be displayed as , together with the countdown from 120s to 0s. Short pressing
it again, you can cancel alarm inhibition; Long pressing the button, the audible alarm
indicator will be displayed as which indicates the audible alarm off. And then long pressing
the button, you can activate the audible alarm again.
Notes:
1. The alarm sound will go on until alarm disappears or is turned off.
2. After silencing the alarm, the corresponding indicator will indicate this.
3. The power low alarm: the corresponding indication lamp will be
fl
ashing with a red frame.
Assignment of priority:
Paramter
SpO
2
PR
/
/
/
/
Yellow
Flashing with yellow
0.5Hz
Di - Di - Di
Di
20 s
Probe off/Finger out
5 s
PR too high/low
Red
Flashing with yellow
1.5Hz
Di- Di – Di ----- Di - Di
3 s
SpO
2
too high/low
High
item
property
Medium
Low
Value
Lamp Frequency
Audiblesound
Alarm cycle
Alarm info
Alarm lamp
- 13 -
ALARM PRIORITY:
There are three-level priorities for selection.
High priority:
indicates the patient is in a dangerous situation.
Medium priority:
indicates the warnings should be paid attention to.
Low priority:
indicates the technical alarm caused by the device itself.
Alarms of the pulse oximeter include technical and physiological alarms. All the three priorities
are divided by built-in module and can not be changed by user.
Warnings!
When an alarm occurs, check patients’ conditions immediately.
●
Check the parameter which is alarming.
●
Check patient’s condition.
●
Search for the source of the alarm.
●
Make the alarm mute if necessary.
●
Check the alarm when no warning.