MedSource MS-74011 Скачать руководство пользователя страница 1

Copyright

Our company owns all rights of this unpublished work and intends to maintain it as a 
con

fi

 dential work. We may also seek to maintain this work as an unpublished copyright. This 

publication is to be used solely for the purpose of reference or operation of our device. No 
part of this work can be disseminated for other purposes.

In the event of inadvertent or deliberate publication, we intend to enforce its rights to this work 
under copyright laws as a published work. Those having access to this work may not copy, 
use, or disclose the information in this work unless expressly authorized by our company.

All information contained in this publication is believed to be correct. We shall not be liable for 
errors contained herein nor for incidental or consequential damages in connection with the 
furnishing, performance, or use of this material. The information this publication refers to is 
protected by copyrights or patents and does not convey any license under the patent rights 
of our company, nor the rights of others. We do not assume any liability arising out of any 
infringements of patents or other rights of third parties.

Content of the manual is subject to change without prior notice.

ALL RIGHTS RESERVED

Bluetooth

®

 and the Bluetooth

®

 Logo are registered trademarks of Bluetooth

®

 SIG, Inc.

Ver2.0  

                        

Issue date: January 18, 2011

Содержание MS-74011

Страница 1: ...k unless expressly authorized by our company All information contained in this publication is believed to be correct We shall not be liable for errors contained herein nor for incidental or consequential damages in connection with the furnishing performance or use of this material The information this publication refers to is protected by copyrights or patents and does not convey any license under...

Страница 2: ... 8 CHAPTER 4 ID Date and Time Setting 9 4 1 Date Time Settings 9 4 2 User ID Setting 9 CHAPTER 5 Take a Measurement 10 CHAPTER 6 Other Settings 12 6 1 Alarm Setting 12 6 2 Data Management 14 6 3 System Setting 16 CHAPTER 7 Maintain and Repair 18 7 1 Maintenance 18 7 2 Safety Checks 18 7 3 Calibration and Verification 19 7 4 Trouble Shooting 20 7 5 Warranty and Repair 20 7 6 Storage and Transportat...

Страница 3: ...MS 7 4 0 1 1 MS 7 4 0 1 1 ...

Страница 4: ... liquid cleaning agents Always disconnect the device from the AC adapter before performing cleaning of maintenance Alarm must be set up according to different situations of individual patient Make sure that chime sound can be activated when alarm function begins to work Cautions Notes Application of this device may influence the measuring accuracy in the background of electromagnetic areas such as...

Страница 5: ...ian This pulse oximeter is designed and tested in compliance with the EMC standard complying with the international standard for the EMC of the electronic medical device IEC 60601 1 2 However because of the proliferation of radio frequency transmitting equipment and other sources of electrical noise in the health care and home environments e g cellular phones mobile two way radios electrical appli...

Страница 6: ...ually decrease the value 5 Menu Select button For entering main menu or confirming the selection setting 6 Alarm inhibit off button Press this button to inhibit alarm sound for 120 seconds And long press the button power off the alarm 7 Return Lock button On the measuring screen it serves as Lock button On the menu and sub menu screen it serves as Return button 8 Power switch Press and hold it for...

Страница 7: ...l measurement The color of the SpO2 value is red when the SpO2 is outside the alarm limits It shows two dashes throughout probe off and finger out conditions 99 SpO2 high alarm limit indicator 90 SpO2 low alarm limit indicator Pulse amplitude bar It indicates the dynamic pulse amplitude and rate As the detected pulse becomes stronger more bars are illumined with each pulse The reverse is true for ...

Страница 8: ...ation Audible alarm on Pulse Beep on The adapter is connected Battery power indication Pulse Beep off Keyboard is unlocked Note The Bluetooth data transmission is reserved Keyboard is locked USB cable is connected USB cable disconnected SpO2 probe is inserted SpO2 probe off Explanation Bluetooth activated Bluetooth deactivated PULSE OXIMETER Alarm inhibition on off Prevent from rain Date of Manufa...

Страница 9: ...critical condition the battery voltage is 3 5V 0 1V at which time the unit shuts down Cautions Be sure to install batteries with correct polarities Only the approved batteries are recommended to be used Do not use batteries not specified for this unit Do not dispose of batteries in fire If battery fluid gets on your skin or clothing rinse with plenty of clean water immediately Remove the batteries...

Страница 10: ...ulse amplitude bar etc Up to 127 patients ID and 72 hour records storage Visual and three level audible alarms battery power low alarm Adjustable pulse beep and backlight Data transfer to PC for review by USB cable Powered by three AA alkaline Ni MH batteries Battery charge function with adapter or charge stand optional Suitable for adult and pediatric greater than 3 years of age users 2 4 Product...

Страница 11: ...nd then press the menu button to enter the main menu refer to the Fig 4 1 2 Press the Down button to select Date and Time item and then press the Menu button to enter the time setup screen refer to Fig 4 2 Pick different sub items to set and press the Select button to highlight it and then using the Up or Down button to adjust the value At last press the Select button to confirm Set other sub item...

Страница 12: ... will be shown Warnings The measurement would not be performed if the following instances come across in operation Shock Low temperature of hand Have taken vascular activity medicine Anemia carboxyhemoglobin methemoglobin methylene blue Indigo carmine Select the suitable probe in terms of type and dimension Plug probe into SpO2 socket on top panel of pulse oximeter Clip the patient finger to the r...

Страница 13: ... or hypothermia There is arterial occlusion proximal to the probe The patient is in cardiac arrest or shock The probe is too tight There is excessive illumination from light sources such as a surgical lamp a bilirubin lamp or sunlight A blood pressure cuff is inflated on the same extremity as the one to which an SpO2 probe is attached Note SpO2 probe should be kept from the light source e g radial...

Страница 14: ...larm setting Also the limits settings of PR are performed as similarly as the SpO2 limits The range of high limit is 31bpm 235bpm And the range of low limit is 30bpm 234bpm 3 Alarm on off In the Alarm Setting screen press the Up or Down button to move the cursor to the Alarm SpO2 or Alarm PR sub item and press the Select button to highlight it Press the Up or Down button to select ON or OFF refer ...

Страница 15: ...ashing with yellow 0 5Hz Di Di Di Di 20 s Probe off Finger out 5 s PR too high low Red Flashing with yellow 1 5Hz Di Di Di Di Di 3 s SpO2 too high low High item property Medium Low Value Lamp Frequency Audiblesound Alarm cycle Alarm info Alarm lamp 13 ALARM PRIORITY There are three level priorities for selection High priority indicates the patient is in a dangerous situation Medium priority indica...

Страница 16: ... for free space when no measurement is being taken Fig 6 3 SpO2 Trend PR Trend Clear Data Data Manage Select Return Data Review 6 2 1 Data Review Pick Data Review sub item and enter by pressing the Select button The screen as in Fig 6 4 will appear Pressing the Select button again a dialog box will pop up refer to Fig 6 5 After selecting an ID number press the Up or Down button to select Delete or...

Страница 17: ...ration is referred to 6 2 2 SpO2 Trend Pick Clear Data sub item and enter by pressing the Select button A dialog box will pop up refer to Fig 6 10 Press the Up or Down button to pick Yes or No and then press the Select button to determine whether delete all the records Note Pay attention to the data deletion as the deleted data cannot be restored Fig 6 6 Fig 6 7 SpO2 Trend SpO2 1 1 Time s Select R...

Страница 18: ...ghtness level 1 7 Bright time ON 15 30 45 60 75 90 105 120 seconds Fig 6 12 Fig 6 13 Blue Tooth Beep Setting Display Mode Keyboard Lock System Setting Select Return Brightness About Default Setup System Setting Select Return Power Auto 6 3 1 Brightness 6 3 2 Bluetooth Reserved 6 3 4 Display mode 6 3 3 Beep setting Note The brightness cannot be set in the charging process Enter the Beep setting scr...

Страница 19: ...t the auto power off function on or off Note The auto power off time cannot be set in the charging process 6 3 7 About Enter the About screen from the System Setting screen the version of software is displayed 6 3 8 Default Setup Enter the Default Setup screen from the System Setting screen and you can restore the default configurations Note The user cannot restore the factory defaults configurati...

Страница 20: ...ies refer to the instructions delivered with the accessories Before every use or after your pulse oximeter has been used for 6 to 12 months or whenever your pulse oximeter is repaired or upgraded a thorough inspection should be performed by qualified service personnel to ensure the reliability Follow these guidelines when inspecting the equipment Make sure that the environment and power supply mee...

Страница 21: ... also take out the batteries after measuring if the pulse oximeter will not be used for a long time Battery maintenance Please take out battery if you will not use the monitor for a long time Please charge the battery fully if you will not use it for a long time Please charge over 8 hours at the first time or may reduce the battery life If any abnormal phenomena occurs you should stop using immedi...

Страница 22: ...n t power on the pulse oximeter Please check the batteries voltage b Probe OFF alarm Please check if the probe was connected with the pulse oximeter correctly If the probe is with extension cable please check if the extension cable is connected with the probe correctly 7 5 Warranty and Repair 7 5 1 Maintenance Method a Maintenance responding time 9 00am 17 30pm Monday to Friday b Service support O...

Страница 23: ... There are liquid leftover and eyewinker on pulse oximeter and lead to short circuit and plugboard failure All the probes and accessories belong to consumption and beyond free change range Such damage of probe caused by mechanical force doesn t belong to free change range Maintenance seal of the pulse oximeter are not opened Not original package lead to the pulse oximeter damage during transportat...

Страница 24: ...oice to standby with the period of guarantee Please describe failure phenomenon in detail and altogether offer pulse oximeter 7 6 Storage and Transportation Storage Storage Temperature 20 C 55 C Relative Humidity 93 no condensation Transportation Transport by airline train or vessel after packing according to request Package We pack the product with the hard bag We put the foam between the inner b...

Страница 25: ...pulse rate value probe off battery exhausted Alarm Mode audible alarm visual alarm and information Alarm Limits range SpO2 70 100 PR 30 235 bpm Default Limits SpO2 High 100 low 90 PR High 100 bpm low 60 bpm SpO2 Display Range 0 100 Measurement Range 70 100 Resolution 1 Accuracy 80 100 2 Digits 70 79 3 Digits 70 unspecified Pulse Rate Display Range 0 254 bpm Measurement Range 30 235 bpm Resolution ...

Страница 26: ...5 x 3 x 1 6 9o u n c e s e x c l u d i n gb a t t e r i e s ...

Страница 27: ... fluctuations IEC 61000 3 3 Compliance Group 1 Group B Class A Complies Electromagnetic environment guidance The model MS 74012 Pulse Oximeter uses RF energy only for its internal function Therefore its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment The model MS 74012 Pulse Oximeter is intended for use in the electromagnetic specified below Th...

Страница 28: ...vironment guidance NOTE UT is the a c mains voltage prior to application of the test level Floors should be wood concrete or ceramic tile If floors are covered with synthetic material the relative humidity should be at least 30 Mains power quality should be that of a typical commercial or hospital environment Mains power quality should be that of a typical commercial or hospital environment Power ...

Страница 29: ...used no closer to any part of the model MD300M Pulse Oximeter including cables than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter Recommended separation distance Where P is the maximum output power rating of the transmitter in Watts W according to the transmitter manufacturer and d is the recommended separation distance in metres m ...

Страница 30: ...plies NOTE2 These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and reflection from structures objects and people a Field strengths from fixed transmitters such as base situation for radio cellular cordless telephones and land mobile radios amateur radio AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy To ...

Страница 31: ...Distributed by MedSource International LLC 5346 Shoreline Dr Mound MN 55364 ...