TROUBLESHOOTING GUIDE 

 
 

PROBLEM 

POSSIBLE CAUSE 

CORRECTIVE ACTION 

 

Nothing happens when the 
button is pressed on  

 

Bed is not plugged in. 

 

Plug the bed’s power cord into an 
electrical outlet. 

the patient hand control 
pendant. 

 

Fuse is blown or your home’s circuit 
breaker is tripped. 

 

 

Check fuse or circuit breaker. 
 

 

 

Motor’s thermal switch was tripped. 

 

Let motor cool down and try again. 

 

 

Hand control is not connected to the 
motor control box. 

 

 

Check to be sure pendant control 
is properly plugged in to the junction box. 

 

 

Power cord, junction box, hand control 
cord or cables are damaged or defective. 

 

 

Call your Medline dealer for service and 
repair. 

 

 

End of stroke is reached. 

 

Operate opposite button. 

 

 

Cords are not properly connected to the 
junction box. 

 

 

Ensure cords are seated securely in the 
appropriate receptacle. 

 

 

Motor will not function after other 
possible causes have been eliminated. 

 

 

Call your Medline dealer for service and 
repair. 

 

The head (or foot) of the 
bed is always higher than  

 

The bed ends were not leveled when the 
bed was assembled. 

 

See instructions for “Leveling the Bed”. 

the other end of the bed. 

 

Two footboards (or headboards) were 
used instead of one of each. Or, 
headboard and footboards are switched 
incorrectly. 

 

 

Headboard is taller than footboard; check 
and exchange. 

 

 

Only headboard (or foot-
board) raises when Hi-Lo 
adjustment is made. 

 

The pin in the middle of the center drive 
shaft has fallen out, allowing the shaft to 
collapse and disengage. 

 

 

Replace the pin and reconnect the 
center drive shaft into the shaft ends.

 

 

 

The center drive shaft ends are not 
properly engaged with the pins at each 
end. 

 

 

Check both ends of the drive shaft for 
proper connection. 
 

 

 

One or more of the pins in the center 
drive shaft have broken. 

 

 

Call your Medline dealer for service and 
repair. 

 

 

Damaged mechanism inside the 
bed boards. 

 

 

Call your Medline dealer for service and 
repair. 

Back or knee section will 
not raise or lower even 
though you turn the crank. 

 

There are internal problems with the 
drive shaft mechanism. 

 

Check to see that the drive mechanism has 
been properly installed. May need to call 
your Medline dealer for service and repair. 

 

 

 

The pull tube is not correctly installed. 

 

Ensure pull tubes are correctly connected.  
Check for binding.  May need to call your 
Medline dealer for service and repair. 

 

Hi-Lo function does not 
work 

 

Manual and semi-electric beds: crank not 
properly installed.   
Full electric bed:  drive shaft 
disconnected. 

 

Check/reinstall drive shaft.  Ensure cord is 
in proper place on junction box.  May need 
to call your Medline dealer for service and 
repair. 

 

Pressing hand control 
buttons results in motion 
other than intended 

 

Motor cable connections to junction box 
are incorrect. 

 

Verify cable connections to the junction 
box are in the correct receptacle. 

 
 
 
 

Содержание MDR107000P

Страница 1: ...MDR107003L Semi Electric Bed Model MDR107002L Manual Bed Model MDR107000P DEALER This manual must be given to the user of the Medline Home Care Bed USER Please read this manual completely for warning...

Страница 2: ...ic 2 Motor 1 Crank Manual 3 Crank Model Number MDR107003L MDR107002L MDR107000P Head Ends Min 28 5 Max 35 75 Min 28 5 Max 35 75 Min 28 5 Max 35 75 Foot Ends Min 26 Max 33 Min 26 Max 33 Min 26 Max 33 F...

Страница 3: ...E AT 1 800 MEDLINE BEFORE USE OTHERWISE SERIOUS BODILY INJURY OR PRODUCT DAMAGE MAY OCCUR A QUALIFIED TECHNICIAN MUST PERFORM ANY REPAIR OR CHANGE TO THIS BED OR SERIOUS INJURY MAY OCCUR TECHNICIANS D...

Страница 4: ...mes 13 Connecting the Spring Fabric 14 Installing the Casters 15 Installing the Headboard and Footboard 16 Patient Hand Control 17 Adjusting Connecting the Head Spring Pull Tube 18 Installing the Cent...

Страница 5: ...5 MEDLITE HOMECARE BED PARTS DIAGRAM Full Electric Bed See Page 7 for List of Terms Head Spring Assembly Foot Spring Assembly...

Страница 6: ...ear Box for Head Board All MDR620025 1 EA 15 Gear Box For Footboard With Handle S M MDR620015 1 EA Gear Box For Footboard No Handle F MDR630035 1 EA 16 Handle Assembly For Hi Low S M MDR620085 1 EA 17...

Страница 7: ...r further instructions Do not use near explosive gases Possible fire hazard when used with oxygen administering equipment other than nasal or masked type Use ONLY mask or nasal type oxygen administeri...

Страница 8: ...head and foot boards BEFORE use Ensure that the pendant and power cords are routed and secured properly so the cords do not become entangled pinched and or severed during operation of the electric bed...

Страница 9: ...nd thickness size or density of the mattress could cause entrapment Use ONLY authorized Medline replacement parts For semi electric and full electric beds always remove the manual crank before perform...

Страница 10: ...rom the junction box DO NOT operate the bed unless the ground wire is attached to frame Weight Limitations The TOTAL weight limit of the Medline Home Care Bed including accessories mattress occupant a...

Страница 11: ...nd Semi Electric Models include motors a junction box and Electrical cord already attached Manual Model includes handle cranks on the frame 1 Hand Pendant Control and Cord Full and Semi Electric Model...

Страница 12: ...sadjustment of motors which may cause frame damage DO NOT attempt to operate the bed controls until assembly of the entire bed is complete as damage to bed components or personal injury may occur Bed...

Страница 13: ...ngle to one another 5 Hook the bottom head spring center mounting latches to the bottom foot spring center mounting rivets see Figure 2a 6 Hook the top head spring center mounting latches to the top f...

Страница 14: ...g up lift the head spring section upward to give slack to the fabric Connect the spring fabric together with the links provided 2 Hook together all the links to the spring fabric 3 Lower the head spri...

Страница 15: ...shaft of the caster into the caster socket of each bed leg Ensure the stem is fully inserted into the socket See Figure 4 4 To prevent excess movement of the bed during assembly lock each of the two l...

Страница 16: ...as close to the head frame as possible 2 Raise the head frame until the rivets on the corner plates of the head frame fit into the corner locks on the headboard The headboard may need to be tilted bac...

Страница 17: ...e hand pendant cable is routed in a way such that it will not become entangled or pinched when the bed is adjusted Do not place the pendant hand control under or between objects This may unintentional...

Страница 18: ...th a telescoping tube Type 1 Please locate which head spring extension tube your bed contains and use the instructions regarding that type of extension tube Cotter Pin Grommet Washer Clevis Pin Headsp...

Страница 19: ...serting the clevis pin into the lift arm slot Reconnect the grommet and washer and secure the cotter pin Figure 9B Type 2 Extension tube attachment 1 To provide up and down movement of the head sectio...

Страница 20: ...ype of bed being used and engage the spring button into the proper positioning hole as follows a For semi electric and manual beds engage the spring button into the first hole position in the outer sh...

Страница 21: ...ents are securely installed BED ASSEMBLY Junction Box Connections 1 To insert cords into the junction box line the cords up with the appropriate receptacle on the junction box and push in until fully...

Страница 22: ...ase the button permit a slight pause and then activate the next operation NEVER operate the bed if the pendant hand control or the power cord is damaged Before placing the bed into use test it by oper...

Страница 23: ...ed Hi Lo Turning the crank clockwise will cause each bed function to raise Turning the crank counterclockwise will cause each bed function to lower Caution After using any hand crank be sure to push t...

Страница 24: ...ot be washed down with a pressure washers or steam cleaner Doing so can result in problems with the electrical system 3 Inspect bed components for any exposed metal To prevent rusting Medline recommen...

Страница 25: ...to ensure there is no damage or excessive wear 8 Inspect all bed motions headspring raises lowers properly foot spring raises lowers properly bed height raises lowers properly 9 Ensure all electrical...

Страница 26: ...headboard or foot board raises when Hi Lo adjustment is made The pin in the middle of the center drive shaft has fallen out allowing the shaft to collapse and disengage Replace the pin and reconnect t...

Страница 27: ...semblies of this Split Spring Home Care Bed These two motors are controlled by a Hand Pendant that is used by the patient or care giver to adjust the bed to either a comfortable or functional position...

Отзывы: