EVENCARE®

 

ProView

Viewing Meter Memory

NOTE:

  If the meter memory is full, adding a new test result will cause the oldest one  

to be deleted.

  Meter memory test results and day averages will not be erased when battery  

is removed.

  The meter will automatically turn off after 2 minutes of non-use.

Содержание EVENCARE ProView

Страница 1: ...BLOOD GLUCOSE MONITORING SYSTEM USER S GUIDE EVENCARE ProView...

Страница 2: ......

Страница 3: ...ed to use and maintain the EVENCARE ProView Blood Glucose Monitoring System Please read this User Manual carefully before you start the test and do a quality control test Confirm the measuring unit on...

Страница 4: ...at 1 866 274 2131 Intended Use The EVENCARE ProView Blood Glucose Monitoring System is intended for the quantitative measurement of glucose in fresh capillary whole blood samples drawn from the finge...

Страница 5: ...al be sure to keep the test strip vial closed tightly at all times For the most accurate results make sure your hands are clean and dry before removing the test strips from their vial Do not touch tes...

Страница 6: ...Keep test strips away from direct sunlight and heat Use this system only for blood glucose test Before using the ProView Blood Glucose Monitoring System please read this manual carefully The meter and...

Страница 7: ...viduals experiencing a hyperglycemic hyperosmolar state with or without ketosis Critically ill patients should not be tested with this blood glucose meter Interferences Acetaminophen salicylates uric...

Страница 8: ...te results Healthcare Professional Information Healthcare professionals performing blood glucose tests with this system on multiple patients must always wear gloves and should follow the infection con...

Страница 9: ...may only be used for testing multiple patients when standard precautions and the manufacturer s disinfection procedures are followed Only auto disabling single use lancing devices should be used with...

Страница 10: ...creen 12 EVENCARE ProView Test Strip 13 EVENCARE ProView Control Solutions 14 2 Meter Setup Inserting the Batteries 15 Setting the Date and Time Beeper and Deleting Results 17 Control Solution Testing...

Страница 11: ...Results Blood Glucose Test Results 35 5 Viewing Meter Memory Viewing Stored Readings and Averages 37 6 Caring For the Meter 40 7 Troubleshooting 47 8 System Specifications 52 9 Product Warranty 53 10...

Страница 12: ...f Kit The EVENCARE ProView Blood Glucose Monitoring System includes the following items EVENCARE ProView Meter User s Guide Carrying Case NOTE ProView Test Strips ProView Control Solutions are require...

Страница 13: ...NCARE ProView Meter BLOOD GLUCOSE M ONITORING SYSTEM EVENCARE ProView Seria l 2 1 GTIN 0 1 10888277409163 M PH4540 TechnicalSupport 1 866 274 2131 M on Fri 8am 5pm CT Display M Button Front Back Test...

Страница 14: ...CARE ProView Blood Glucose Monitoring System Meter Display Screen Low Battery Warning Stored Test Results Test Results Unit of Measure Temperature Error Beeping Sound Muted Ready to Test Control Solut...

Страница 15: ...lood Glucose Test Strips with the EVENCARE ProView Meter Other brands of test strips will not work with the meter Keep the test strip vial capped tightly and away from sunlight at all times Replace th...

Страница 16: ...olution Testing EVENCARE ProView Glucose Control Solution helps to validate the performance of the system It is available at two levels Level 2 and Level 3 Run the control test to make sure the meter...

Страница 17: ...eries CR2032 Step 1 Remove the battery compartment cover Step 2 If there are old batteries in the compartment take them out Insert new batteries with the plus side facing up Step 3 Replace the compart...

Страница 18: ...16 EVENCARE ProView Meter Setup IMPORTANT Dispose of used batteries according to local regulations NOTE Stored test results will not be erased when the batteries are changed...

Страница 19: ...e date and time after you change the batteries Step 1 Turn on the meter Press and hold the M button for 3 seconds until you see SET on the meter screen to turn the meter on and enter set mode Step 2 S...

Страница 20: ...Press the C button to change select the desired time format 12 hour or 24 hour Press the M button to confirm the selection and advance to set the hour Step 5 Set the hour The hour will flash Press the...

Страница 21: ...ter at any time Continue to Step 8 to set the beeper Step 8 Set the beeper You have the option to turn the beeper on or off You will see ON or OFF flashing on the screen and the word BEEP at the botto...

Страница 22: ...rpose of the control solution testing is to validate that the EVENCARE ProView Meter is working properly with the test strips Control solutions are not included in the kit Please contact your local su...

Страница 23: ...conditions Practicing your testing technique Control Solution Testing Materials needed EVENCARE ProView Meter EVENCARE ProView Glucose Control Solutions Level 2 or Level 3 A new EVENCARE ProView Blood...

Страница 24: ...Step 1 Insert a test strip into the meter with the arrow facing up Make sure to firmly and completely insert the test strip into the meter The meter will turn on automatically IMPORTANT Do not use te...

Страница 25: ...n the screen shake the bottle of control solution and then squeeze a drop of the control solution onto a clean dry non absorbent surface such as table top Bring the test strip gently to the drop of co...

Страница 26: ...he control solution bottle symbol will appear on the display with the result Step 5 Compare the result on the screen to the control solution range printed on the test strip bottle NOTE Mark all contro...

Страница 27: ...sure glucose control solution and test strips have not expired Repeat the test with a new test strip Control solution test results will be stored in the meter memory with a control solution bottle sym...

Страница 28: ...low the manufacturer s instructions for how the lancing device should be used IMPORTANT Do not leave the lancing device after use be sure to dispose of properly Only an auto disabling single use lanci...

Страница 29: ...s between patients and healthcare professionals Please follow the practice guideline below Biosafety in Microbiological and Biomedical Laboratories BMBL found at http www cdc gov biosafety publication...

Страница 30: ...he fingertip palm and forearm 1 alternative site test results may be significantly different than results from the fingertip under certain conditions Alternative site testing should only be performed...

Страница 31: ...ondition Patient s routine glucose results are often fluctuating If the patients alternative site testing do not agree with how they feel IMPORTANT Never change any treatment without first consulting...

Страница 32: ...g a Blood Glucose Test Step 1 Insert a test strip into the meter The meter will turn on automatically Make sure it is inserted completely without bending the test strip IMPORTANT A new pair of clean g...

Страница 33: ...nufacturer s instructions for how the lancing device should be used IMPORTANT Gently massage the hand and finger toward the puncture site to form a drop of blood Do not milk or squeeze around the punc...

Страница 34: ...release pressure to the lancing device for a few seconds while keeping the lancing device in constant contact with the skin When a drop of blood is formed lift the lancing device straight up without...

Страница 35: ...he blood drop onto the sampling end DO NOT add or apply a second drop of blood This may cause a false result When there is an insufficient blood sample the strip icon and the message Er 4 will appear...

Страница 36: ...5 Eject the test strip into a proper waste container IMPORTANT If you see HI or LO displayed the patient s blood glucose level may be above 600 mg dL or below 20 mg dL Repeat the blood glucose test I...

Страница 37: ...mg dL before meals Less than 140 mg dL 2 hours after meals If the meter displays a result that is HI or LO OR a patient gets a result that is more than 250 mg dL or less than 50 mg dL AND the patient...

Страница 38: ...duals or patients in shock Inaccurate low results may occur for individuals experiencing a hyperglycemic hyperosmolar state with or without ketosis Critically ill patients should not be tested with bl...

Страница 39: ...Meter automatically stores up to 300 glucose test results together with the date and time The meter also provide a 14 day average to help track a patient s blood glucose trend The time and date must b...

Страница 40: ...next and so on Press and hold the C button to flash through the results in the meter Step 4 After the oldest result in meter memory continue to press the C button to view the day averages The meter w...

Страница 41: ...If the meter memory is full adding a new test result will cause the oldest one to be deleted Meter memory test results and day averages will not be erased when battery is removed The meter will autom...

Страница 42: ...the meter DO NOT use glass cleaners or household cleaners or corrosive liquid e g Benzene Acetone on the meter DO NOT try to get moisture on the test strip test strip slot or serial port DO NOT take t...

Страница 43: ...the auto disabling safety model marketed for single use only Glucose meters used in a clinical setting for testing multiple persons must be cleaned and disinfected between patients Used gloves should...

Страница 44: ...splay Corroding of the plastic housing Corroding of the buttons Malfunction of any meter buttons If these signs of deterioration are noted stop using the meter and call Medline Technical Service at 1...

Страница 45: ...with a new CaviWipes towelette or an EPA registered disinfecting wipe The cleaning process does not reduce the risk for transmission of infectious diseases Disinfection Instructions The meter must be...

Страница 46: ...t www medexsupply com This product can also be found through large online retail distribution sites such as www amazon com www ebay com etc Cleaning Instructions Step 1 Wash hands with soap and water...

Страница 47: ...trip port and serial port Do not wrap the meter in a wipe Step 5 If using the CaviWipes towelette allow to remain wet for two minutes For other EPA registered disinfecting wipes allow the surface of t...

Страница 48: ...t 50 F to 104 F for about twenty 20 minutes Avoid leaving the meter in extremely hot or cold places such as near a heat source or in an extremely hot or cold car Do not store or use the meter or test...

Страница 49: ...the test strip and insert a new one Apply blood to the test strip after the blood symbol appears on the screen Troubleshooting 7 Troubleshooting This section details display screen messages and error...

Страница 50: ...es or until the meter has reached the proper range and retest Do not raise or lower the meter temperature artificially There is not enough blood in the strip to perform a test The wrong strip has been...

Страница 51: ...he patient s fingertip with a new test strip If the result is still HI contact the patient s physician or healthcare provider Glucose test result is lower than 20 mg dL Wash and dry your hands and rep...

Страница 52: ...ied to the test strip correctly Make sure the test strip is inserted correctly and repeat the test See if the test strip is covered in debris or lint Also make sure the test strip has not been out of...

Страница 53: ...he batteries are inserted correctly with the sign up and facing the right direction Place the meter and test strip in a temperature environment of 50 F 104 F You should use it after about 20 minutes i...

Страница 54: ...it Range 30 55 Display Type LCD screen Memory 300 test results Result Averaging 14 day averaging Battery 2 x 3V CR2032 Lithium coin cell battery Automatic Power Off After 2 minutes of non use Operatin...

Страница 55: ...ect unauthorized repair or modification This warranty is void and of no force and effect in the event of product misuse improper use abuse neglect unauthorized repair or modification This warranty spe...

Страница 56: ...ome into contact with blood Used meters therefore carry a risk of infection Please dispose of your used meter after removing the batteries according to the regulations applicable in your area For info...

Страница 57: ...hat the device will perform as intended The equipment complies with the emission and immunity requirements described in EN 61326 2 6 The electromagnetic environment should be evaluated prior to operat...

Страница 58: ...es Inc Medline and EVENCARE are registered trademarks of Medline Industries Inc Manufactured in China for Medline Industries Inc Three Lakes Drive Northfield IL 60093 USA Technical Support 1 866 274 2...

Страница 59: ......

Страница 60: ......

Отзывы: