14
OPERA
TOR
'S MANU
AL
WHAT YOU SEE
WHAT IT MEANS
WHAT YOU SHOULD DO
Battery is low.
Change the battery according
to the User’s Guide.
Glucose test result is
higher than 600 mg/dL.
Wash and dry your hands
and repeat the test on the
patient’s fingertip with a
new test strip. If the result
is still “HI,” contact the
patient’s physician or
healthcare provider.
Glucose test result is
lower than 20 mg/dL.
Wash and dry your hands
and repeat the test on the
patient’s fingertip with a
new test strip. If the result
is still “LO,” contact the
patient’s physician or
healthcare provider.
There are no results
stored in the meter.
Check if the date and time
are set up correctly.
WARRANTY
The EVENCARE G3 Blood Glucose Meter is warranted to be free of defects in workmanship and materials
for a period of two (2) years from the date of purchase. Medline’s liability for warranty claims is limited
to repair or replacement, and in no event shall Medline be liable for indirect or consequential damages,
or for any loss arising from misuse, improper use, abuse, neglect, unauthorized repair or modification.
This warranty is void and of no force and effect in the event of product misuse, improper
use, abuse, neglect, unauthorized repair or modification. This warranty specifically
excludes EVENCARE G3 Test Strips and EVENCARE G3 Control Solution.
THIS WARRANTY IS THE SOLE AND EXCLUSIVE WARRANTY TO THE EXCLUSION OF ALL OTHER
WARRANTIES, EXPRESS OR IMPLIED, ALL OF WHICH ARE WAIVED AND DISCLAIMED.
All warranty claims must be directed to Medline’s Technical Service Center at 1-866-274-2131.
This warranty is extended only to the original purchaser of the meter.