SELF TEST—ANSWER KEY
1. True or
False
:
EVENCARE G3 needs to be calibrated/coded.
2. What is the purpose of testing with control solutions?
To validate that the EVENCARE G3 meter is working properly with the test strips.
3. When performing a control solution test, you should:
a. Apply the control solution direct from the bottle to the test strip
b. Squeeze a drop of control solution to a flat surface, and touch the test strip to the control solution
c. Apply control solution to the test strip before inserting it into the meter
4.
True
or False:
You need to mark a control test as a control test, so that it shows as a control test in the memory of the meter.
5.
True
or False:
For vial test strips, the vial cap should be closed immediately after removing a test strip from the vial to avoid
any degrading or contamination of the test strips.
6. How long is the test time on the EVENCARE G3 meter?
6 seconds
7. Where can you find information to troubleshoot error messages?
a. User’s Guide
b. EVENCARE Technical Support Line
c. All of the above
8. What type of battery is used in the EVENCARE G3 meter?
One 3V CR2032
9. What is the procedure for disinfecting the EVENCARE G3 meter?
Clean the meter with a disinfecting wipe.
Wipe all external areas of the meter including both front
and back surfaces until visibly wet. Allow the surface of the meter to remain wet at room temperature
for the contact time/kill time listed on the canister. Then, wipe meter dry or allow to air dry.