25
EvenCare
®
G3
Blood Glucose System
Performing a Control Solution Test
We reserve the right to correct any errors.
©2014 Medline Industries, Inc., One Medline Place, Mundelein, IL 60060 | 1-800-MEDLINE | www.medline.com
EvenCare and Medline are registered trademarks of Medline Industries, Inc.
MKT1442259 / LIT154 / PDF / 10
step
1
step
4
step
2
step
5
step
3
step
6
Insert test strip to
turn on meter.
Compare result to the low
or high range printed on
test strip vial.
Invert and gently shake
the control solution
bottle three times.
Apply drop to top of
test strip vial.
Eject and discard the
used test strip. The meter
will automatically turn off.
Wipe clean the top of the
strip vial.
Press or once to
enter control solution
mode for low or high.
The “ctl” icon will display
in the lower left corner of
the LCD, indicating the
meter is in control mode.
STEP 1
STEP 2
STEP 5
STEP 3
STEP 4
STEP 6
STEP 1
STEP 2
STEP 5
STEP 3
STEP 4
STEP 6
STEP 1
STEP 2
STEP 5
STEP 3
STEP 4
STEP 6
STEP 1
STEP 2
STEP 5
STEP 3
STEP 4
STEP 6
STEP 1
STEP 2
STEP 5
STEP 3
STEP 4
STEP 6
STEP 1
STEP 2
STEP 5
STEP 3
STEP 4
STEP 6
CAUTION:
Failure to follow instruction may cause out of range results.
NOTE:
• Write down the date when the control solution is first opened. Discard three months after the date of the first opening.
• For more information on performing a control solution test, see the EvenCare G3 User’s Guide or call Customer Support
at
1-866-274-2131
.
Bring meter and strip to the
droplet. The meter will beep
and display count down bars
after enough control solution
is drawn into the test strip.
46-76
184-250
46-76
184-250
46-76
184-250
46-76
184-250