8
CONTROLS
(CONT.)
WATER SIDE DISPLAYS
29.
SET VOLUME,
displays the vol-
ume of water to be delivered
over the selected Time Interval
or at the end of the dose limit.
Select from 0–500 mL in 1 mL
increments.
30.
TIME INTERVAL,
displays the
time interval (2 to 12 hours in 1
hour increments) over which the
selected water volume is deliv-
ered. When doSE is displayed,
indicates Programmable Flush
is activated. When
nOnE
is dis-
played, all water side settings
are deactivated.
31.
VOLUME DELIVERED,
displays
amount of selected water vol-
ume that has been delivered.
At the conclusion of the Time
Interval selected, the Volume
Delivered turns to 0 and a new
accumulation begins.
32.
ACCUMULATED VOLUME,
dis-
plays water volume delivered
over multiple Time Intervals.
FEEDING SIDE DISPLAYS
24.
FEEDING RATE,
displays the
feeding infusion rate. Select from
1–295 mL/hr in 1 mL increments.
25.
VOLUME DELIVERED,
displays
full strength or diluted feeding
volume delivered.
26.
DOSE LIMIT,
displays dose limit.
Select from 1–2000 mL in 1 mL
increments. Dose is completed
when Volume Delivered equals
the Dose Limit set. Refer to
Dose Delivered under Alarms/
Indicators. If
nOnE
is displayed,
this feature is deactivated.
27.
ACCUMULATED VOLUME,
dis-
plays full strength or diluted
feeding volume delivered over
multiple feedings.
28.
STRENGTH,
(Feeding Dilution)
displays feeding strength to be
delivered. Select from
FS
(Full
Strength),
3
/
4
,
2
/
3
,
1
/
2
,
1
/
3
,
1
/
4
strength, or
H
2
0
(water only).