
18
TROUBLESHOOTING
EntraFlo H20™ Nutrition Delivery Systems are reliable electromechanical devices. As
with any mechanical device, problems may occur. The following tips will help you cor-
rect the minor in-service problems that may arise.
POSSIBLE CAUSE
• Battery charge is below
performance level
• Feeding or water container
is empty
• Occlusion is restricting flow
• Drip chamber not properly
placed in pump
• Drip chamber walls coated with
feeding drops or
condensation in sensor path
• Flow sensors housing need
cleaning
• Pump is on hold for more than
2 1/2 minutes
• Pump has delivered
preset volume
• Door not closed and latched
• Wrong set is used
• Set is not properly in place
SOLUTION
• Plug cord into AC outlet
• Replace or refill feeding or
water container
• Identify point of occlusion and
free up
- Pump Set
• Kinked tubing
• Blocked tubing
• Closed slide clamp
- Feeding tube
- Feeding container
• Assemble properly
• Manipulate chamber to reset
drops or condensation in
sensor path
• Clean with cotton swab damp-
ened with isopropyl alcohol
• Press RUN/PAUSE button
• Reset Volume Delivered and
start a new dose
• Examine tubing for proper
placement, close door
•
Load an appropriate set
• Assemble tubing properly
around pump rotor. Close and
latch door.
TROUBLESHOOTING GUIDE
SITUATION
• LOW BATTERY
alarm activated
• (
LOW BATTERY
)
illuminated
• (
FEEd Err
)
alarm activated
• (
HoLd Err
) Pump is
left on hold
• (
doSE dEL
) Dose
Complete
• (
no SEt
) alarm
activated
Содержание EntraFloH2O
Страница 1: ...ENT199255 Operating Instructions Guide d utilisation Instrucciones de operación ...
Страница 3: ......
Страница 67: ......