EN
13
FLOW/INFUSION RATE:
1-295 mL/hr in 1 mL increments
DOSE LIMIT:
0-9995 mL in 5 mL increments
ACCURACY:
±10% of selected flow rate
OCCLUSION PRESSURE:
Does not exceed 15 p.s.i.
BATTERY:
• Sealed lead acid
• Rechargeable
• Automatically recharges when connected to
an AC outlet
• Recharge time is 12 hours from complete
discharge to full recharge
• Operating time is 8 hours at 100 mL/hr
(Charge unit fully every 30 days)
The backup battery is designed to provide
temporary power to the pump if AC outlet
power is interrupted. Frequent, extended
operation of the pump on the backup battery
will shorten battery life.
DIMENSIONS:
7.4” High x 9.5” Wide x 5.3” Deep
WEIGHT:
5.7 Pounds
CASE MATERIAL:
Fire Resistant Plastic
POWER REQUIREMENT:
120V ~, 187mA, 60Hz
Serial Number less than 700000:
Serial Numbers 850000 or greater:
All Serial Numbers:
Protected Against Vertically Falling Water
Drops
IPX1
Type of Protection Against Electrical Shock
Class I
Type of Protection Against Electrical Shock
Class II
Protection Against Access to Hazardous Parts
with a Finger and Vertically Falling Water Drops
IP21
Expected Battery Life
2 years (200 full charges)
Expected Life of Device
5 years
When integrity of external Mains is in question - use battery power only
Mode of Operation
Continuous
Operating Temperature
5°C - 40°C (41°F - 104°F)
95% R.H. non-condensing
Storage Temperature
0°C - 50°C (32°F - 122°F)
10% - 95% R.H. non-condensing
Sound Pressure Level
Does not exceed 70dB(A)
Degree Of Protection Against Electrical Shock
TYPE BF (Administration Sets are non-
conductive)
Operating Barometric Pressure
700 hPa (10.2 p.s.i.) - 1060 hPa (15.4 p.s.i).
Pump components are not affected by pres-
sure conditions.