PCF Pro User Manual
DOC01851
Page 41 of 74
Testing Operation of the Safety Devices
To check operation of the carrier lift:
1. Press and hold to fully open the
carrier lift and confirm the display
shows:
2. Press and hold to fully close the
carrier lift and confirm the display
shows:
3. If the carrier lift system encounters a
problem it will be shown on the
display:
4. If a carrier limit fault message is
shown please contact Medisafe to
resolve the fault.
To check operation of the safety edge:
1. While moving the carrier press the
safety edge and confirm the carrier lift
stops and the display shows:
2. Press
Reset Safety Edge
on the
display and confirm the carrier lift can
be moved again.
To check operation of the emergency stop:
1. While moving the carrier lift press the
emergency stop and confirm the
carrier lift stops and the display shows:
2. Pull out the emergency stop button
then press
OK
on the display, and
confirm the carrier lift can be moved
again.
Содержание PCF Pro
Страница 1: ...DOC01851 PCF Pro User Manual...
Страница 7: ...DOC01851 Page 7 of 74 1 Introduction 1 Introduction User Manual...
Страница 19: ...DOC01851 Page 19 of 74 2 Installation 2 Installation User Manual...
Страница 26: ...PCF Pro User Manual DOC01851 Page 26 of 74 Installation Drawings Rear View Showing Services...
Страница 27: ...PCF Pro User Manual DOC01851 Page 27 of 74 Front View...
Страница 28: ...PCF Pro User Manual DOC01851 Page 28 of 74 Left Side View with Lid Closed showing Drain Connection...
Страница 29: ...PCF Pro User Manual DOC01851 Page 29 of 74 Left Side View Lid Open...
Страница 31: ...DOC01851 Page 31 of 74 3 Operation 3 Operation User Manual...
Страница 55: ...DOC01851 Page 55 of 74 4 Troubleshooting 4 Troubleshooting User Manual...
Страница 59: ...DOC01851 Page 59 of 74 5 Technical 5 Technical User Manual...
Страница 65: ...DOC01851 Page 65 of 74 6 Appendices 6 Appendices User Manual...