PCF Pro User Manual
DOC01851
Page 14 of 74
Caution – Possible Equipment Damage
Caution
•
Never place items or person(s) on the lid, this is particularly important while
operating the carrier lift.
•
Ensure the thumb ball latches in the down position before moving the carrier
lift.
•
Never place items on the machine which could obstruct the lid from closing.
•
Never attempt to obstruct the lid from closing.
•
All baskets must be fitted when running a cycle.
•
Do not place items directly into the wash tank or allow them to make contact
with the tank sides.
•
Never scratch chamber walls. This can increase cavitation erosion.
•
Do not load basket with instruments pointing at the lid.
•
Do not overload the basket, maximum load of 10kg (22lbs) per basket.
•
Do not allow Luer pipes to protrude from the basket.
•
Always use a Full DIN Tray Standard to hold instruments (M20044).
•
When choosing a detergent, select one with a low chloride content.
Detergents with a high chloride content can be harmful to stainless steel.
•
Repeated fuse blows or circuit breaker trips indicate a short circuit or overload.
Have a fully trained technician locate and correct situation.
•
Use nonabrasive cleaners when cleaning unit. Follow directions on container
and rub in a back-and forth motion, in same direction as surface grain.
Abrasive cleaners damage stainless steel. Cleaner rubbed in a circular motion
or applied with a wire brush or steel wool will scratch and dull stainless steel.
•
Avoid running equipment into the Control Panel or Touch Screen to prevent
damage to controls.
•
When removing baskets always tilt down to ensuring any retained water drips
into the tank.
Caution – Chemical
Caution
•
Chemical bottles must be connected to the correct port. Otherwise, wash
efficiency and/or rinsing will be compromised. Ensure that the chemical type
matches the port shown on the display.
•
Check chemical bottles are in date before use
Caution – Instrument Care
Caution
•
Always confirm compatibility with the instrument manufacturer.
•
Always follow the manufacturer’s instructions if processing instruments with
special characteristics e.g. fibre optics or rigid optics.
•
Whether pre or post processing instruments, always follow the manufacturer’s
instructions.
•
Always attach tubing to a suitable size distal connector for instruments being
cleaned or to a quick connect port if the instrument has one.
•
Always open box jointed instruments.
•
Always ensure taps are open on cannulated instruments.
•
Always dissemble instruments that are designed to be taken apart.
•
Visually inspect instruments after reprocessing
•
If any instruments became detached from the Medisafe connection tubing
during processing, they should be reprocessed after ensuring that the Luer
connection is firmly seated.
Содержание PCF Pro
Страница 1: ...DOC01851 PCF Pro User Manual...
Страница 7: ...DOC01851 Page 7 of 74 1 Introduction 1 Introduction User Manual...
Страница 19: ...DOC01851 Page 19 of 74 2 Installation 2 Installation User Manual...
Страница 26: ...PCF Pro User Manual DOC01851 Page 26 of 74 Installation Drawings Rear View Showing Services...
Страница 27: ...PCF Pro User Manual DOC01851 Page 27 of 74 Front View...
Страница 28: ...PCF Pro User Manual DOC01851 Page 28 of 74 Left Side View with Lid Closed showing Drain Connection...
Страница 29: ...PCF Pro User Manual DOC01851 Page 29 of 74 Left Side View Lid Open...
Страница 31: ...DOC01851 Page 31 of 74 3 Operation 3 Operation User Manual...
Страница 55: ...DOC01851 Page 55 of 74 4 Troubleshooting 4 Troubleshooting User Manual...
Страница 59: ...DOC01851 Page 59 of 74 5 Technical 5 Technical User Manual...
Страница 65: ...DOC01851 Page 65 of 74 6 Appendices 6 Appendices User Manual...