Connecting the PCF Pro
Note
: Medisafe (UK) Ltd. recommends this product be installed & commissioned by Medisafe (UK)
Ltd (or suitably qualified personnel).
1. Unpack the machine and position it within
1m of the mains power supply, drain, and
mains water supplies.
2. Identify the drain hose supplied with the
unit. Connect the hose to the drain outlet
underneath the right hand side of the unit.
3. Secure the drain hose to the machine
using the supplied jubilee clip, and cut to
length. Refer to the drain connection
diagram in section 2-9-3 for connection
options.
4. Connect the machine drain to the site
drain.
5. Connect the compressed air hose to the
8mm connector on the lower rear of the
unit.
6. Identify the blue cold-water hose supplied
with the unit. Ensure the seal in the hose
end is in place and screw the elbow
connector onto the cold-water inlet at the
lower rear of the unit.
7. Connect the straight end of the hose to a
suitable cold water supply.
8. Identify a stainless steel water hose
supplied with the unit. Ensure the seal in
the hose end is in place and screw the
elbow connector onto the cold-water inlet
at the lower rear of the unit.
9. Connect the straight end of the hose to a
suitable hot water supply.
10. Repeat for the RO water if this has been
made available on site.
11. Turn on the water supplies and check for
leaks.
12. Fit the shorter of the kick plates to the
machines side, ensuring the nut insert
faces the front of the machine.
13. Screw the side kick plates to the machine,
and fit the rubber hole covers.
14. Move the machine into its final position,
ensuring the castors are aligned front to
back, and apply the front castor brakes
15. Fit the front kick plate over the load switch,
ensuing the switch operates. Screw the
front kick plate to the machine, and fit the
rubber hole covers.
16. Connect the machine to the electrical
mains supply (see section 2-7)
17. Apply power to the machine and open the
lid (see section 3-7-6)
18. Fit the tank filter ensuring it is inserted flat
into the tank recess.
19. Check carrier lift, emergency stop and
safety edge operation (see section 3-7-10)
20. Carry out the commissioning procedure
defined within the PCF Pro Service
Manual.
Содержание PCF Pro
Страница 1: ...DOC01851 PCF Pro User Manual...
Страница 7: ...DOC01851 Page 7 of 74 1 Introduction 1 Introduction User Manual...
Страница 19: ...DOC01851 Page 19 of 74 2 Installation 2 Installation User Manual...
Страница 26: ...PCF Pro User Manual DOC01851 Page 26 of 74 Installation Drawings Rear View Showing Services...
Страница 27: ...PCF Pro User Manual DOC01851 Page 27 of 74 Front View...
Страница 28: ...PCF Pro User Manual DOC01851 Page 28 of 74 Left Side View with Lid Closed showing Drain Connection...
Страница 29: ...PCF Pro User Manual DOC01851 Page 29 of 74 Left Side View Lid Open...
Страница 31: ...DOC01851 Page 31 of 74 3 Operation 3 Operation User Manual...
Страница 55: ...DOC01851 Page 55 of 74 4 Troubleshooting 4 Troubleshooting User Manual...
Страница 59: ...DOC01851 Page 59 of 74 5 Technical 5 Technical User Manual...
Страница 65: ...DOC01851 Page 65 of 74 6 Appendices 6 Appendices User Manual...