Service Manual - medifa 8000
30
To replace the castors jack up the table.
Place two squared timber underneath the column.
Do as follows :
The gear shift linkage of the radioscopy operation table can be dismantled by unscrewing the six screws.
As soon as these screws are loosened the whole gear shift linkage can be folded.
Do as follows:
First push the fixing brake pedal directly to the hub.
Then push the connection pipe to the pedal.
Attention: When you push the connection pipe to the pedal, it could happen that the short
piece from the gear shift linkage fall close!
Following you have to remove the short piece of the hexagonal bar. After that you can remove the whole gear
shift linkage bit by bit.
XIV. Replacement of the Castors
(Fixed Roller, Back)
fig. 95:dismantle the driving mechanism
fig. 96: dismantle the driving mechanism
fig. 97+98 dismantle the driving mechanism