Service Manual - medifa 8000
Finally, separate the wire of the way measurement system
( lateral shifting) .
Remove the two screws and open the way measurement
cap. You will now see the plug. Simply put it off.
After separating the way measurement system from the cylin-
der and cutting off the connection, simply screw off the holder
of the driving pin.
Remove the two fastening screws and mount the holder di-
rectly to the driving pin of the new system.
Note:
Should the cable be faulty,too, retrace it back to the connec-
tion and replace it.
See the diagram for the correct placement of the cables in
the breaker box.
22
fig. 71+72: disconnect way measurement
system
fig. 73: loosen holder of driving pin