![MEDICSIGHT MEDICCO2LON Скачать руководство пользователя страница 28](http://html.mh-extra.com/html/medicsight/medicco2lon/medicco2lon_user-manual_1761096028.webp)
28
Copyright © 2020 Ultrasound Technologies Ltd
28
6. MAINTENANCE
Pressure Display
is red and alarm
sounds
• Actual pressure is
50mmHg or higher
• Gas delivery line
blocked
1. Unit will automatically vent to target pressure
within several seconds. Thereafter, alarm will
silence.
2. Should pressure remain above target pressure,
power unit down and disconnect Administration
Set from unit.
3. Check patient connection and re-establish as
required.
4. Should checking 1), 2) and 3) fail to rectify
the problem, call distributor representative for
service.
Volume Display and
Volume Reset icons
do not operate as
specified
• Internal fault or
failure
1. Power unit OFF / ON.
2. Set Pressure and/or Flow greater than zero.
3. Should checking 1) and 2) fail to rectify the
problem, call distributor representative for
service.
Start/Resume/
Stop icon does not
operate as specified
• Internal fault or
failure
1. Power unit OFF / ON.
2. Should checking 1) fail to rectify the problem,
contact the distributor representative.
SD card error/SD
card full indicator
displayed
• SD card corrupted
or full.
1. Power unit OFF and Eject the SD card.
2. Replace with a new SD card (SDSC cards only,
available from distributor).
3. Power unit On.
3. Should checking 1) and 2) fail to rectify the
problem, contact the distributor representative.
6.2
Cleaning and Sterilisation Precautions
• Follow universal precautions for protective apparel when handling and cleaning
contaminated instruments.
• Examine the device carefully for any damage before every use. Do not use if
there is any physical degradation, fraying or other damage to the cables, tubes,
hoses or connectors.
• The Administration Set is supplied non-sterile.
• Administration sets are single use and should be disposed of in accordance with
local hospital policy and with regard to all applicable regulations, including but
without limitation to, those pertaining to human health & safety and care of the