To connect PIN Index valve cylinders:
1. Close the regulator by rotating the valve handle counter-clockwise.
2. Align the adaptor yoke on the regulator to the PIN Index connection on the
cylinder.
3. Attach the connection tube supplied to the outlet connection and attach the
other end of the connection tube to the MedicCO
2
LON device.
4. Slowly open the gas cylinder valve. A leaking valve can be identified by listening
for any seepage sound. A leaking valve should be reported immediately as it can
be a serious safety issue.
4.1.5 Gas Supply Indicator
The main screen displays a gas supply indicator. The levels displayed are a guide to
determine whether there is sufficient gas in the CO
2
cylinder to perform a procedure.
Cylinder contains sufficient CO
2
for
multiple procedures.
Unit will begin procedure and supply
CO
2
as required.
Cylinder CO
2
level is low, may not
complete multiple procedures.
Unit will supply CO
2
as required.
Cylinder CO
2
level too low to complete
procedure, change cylinder.
Unit will not begin procedure or supply
CO
2
.
Cylinder is empty or supply not
connected to unit. Check cylinder
connection or change cylinder.
Unit will not begin procedure or supply
CO
2
.
ATTENTION:
The unit will not function if the gas supply indicator is empty.
WARNING:
Do not allow liquid CO
2
to enter the unit. This can be prevented by
ensuring the CO
2
cylinder remains upright at all times.
4. INSTALLATION
and SETUP