17
6. Alarm and
Monitoring
Funcitons
A built-in microcontroler in the
ECO 300
allo-
ws for a permanent monitoring of the most im-
portant parameters.
●
unhindered oxygen flow
●
switching of the magnet valve
●
monitoring of breath impulses
●
monitoring against hypoventilation
●
battery monitoring
When the oxygen flow from the bottle to the
Precise 3000
is impeded an acoutic signal is set
off at every inhalation and
NO O
2
is indicated in
the display. In that case check the spiral hose
connections for correct setting or open the valve,
resp. change the bottle in case it is empty.
If the
Precise 3000
is working correctly the
switching of the magnet valve at inhalations is in-
dicated by flashing of a * in the upper right hand
corner of the display.
When the device is in operation mode and no
breathing takes place for one minute
CHECK
CANULA
is indicated in the display and an
acoustic signal is set off for one minute, interrup-
ted for one minute and set off again for one minu-
te and then the
Precise 3000
turns off automa-
tically storing the current settings. Closing the
bottle valve is part of the actions necessary to end
the treatment.
When the battery power has become too low
CHANGE BATTERY
appears in the display.
In this case change the battery when changing the
bottle but not later than five hours after the indi-
cation in the display .
If the battery power is too low the
Precise
3000
cannot be used.
In case of Précise 3000 failure (battery empty,
technical defect of the Précise 3000 etc.) the nasal
canula can be connected to the spiral hose with
the help of an adapter. A basic alimentation of 2.8
l/min for the user can be ensured in this way.
Operation keys
I/O
Turn on, turn off the device
<
Reduce O
2
flow by 0.5 l/min (mini-
mum 0.5 l/min)
>
Increase O
2
flow by 0.5 l/min (ma-
ximum 0.5 l/min)
Key combination
< >
(see 3.3. operation modes)
Indication in the display
Indications in the upper line
Operation mode
MANUAL
or
AUTOMATIC
Switching of the magnetic valve * (in the right
corner)
Indications in the lower line
Current amount of O
2
0 0.5 1 1.5 2 2.5 3 3.5 4 4.5 5 5.5 6 6.5 7 7.5 8 l/min
In case of alarm the display changes from the in-
dication of the current amount of O
2
to the alarm
message.
Alarm messages:
CHECK CANULA
NO O
2
CHANGE BATTERY
Содержание PRECISE 3000
Страница 11: ...Technical Manual and Instructions for Use PRECISE 3000 Electronic gas conserver ...
Страница 19: ...Description de l appareil PRECISE 3000 Système économiseur de gaz à contrôle électronique ...
Страница 28: ......
Страница 29: ...Technische handleidingen Gebruiksinstructies PRECISE 3000 ...
Страница 38: ...38 medicare homecare GmbH Hoherodskopf Str 22 D 35327 Ulrichstein ...
Страница 39: ......
Страница 40: ...40 ...