CT/ MRI compatible MF-System
IFU GA-008-02
Revision
01
Page
Seite
24
von
37
7. Adjust the connecting arm.
8. Screw the swivel adapter to the connecting
arm.
9. To have additional security and stability, the
screws should be tightened with these tools.
10. Test this assembly to see if it is stable by
pulling/pushing/turning. If not, check if screws
are not tightened properly and if the clamping
adapter clamps properly.
(If skull clamp is used proceed to page 27, for
headrest horseshoe see page 29.)