background image

 

38 

Step 3 

 

“Press the flashing button now.” {Repeat twice} 

 

“Shock delivered.” 

 
When the shock button is not pressed, the following voice prompt will be emitted: 

 

“Press the flashing button now.” {Repeat twice} 

 

“Shock button not pressed.” 

 
When  the  patients’  ECG  rhythm  is  changed  to  a  non-shockable  rhythm,  when  the 
pads  are  disconnected  or  when  the  impedance  or  mode  is  changed,  the  following 
voice prompt will be emitted: 

 

“Shock cancelled.” 

 

 

Figure 21. HeartOn A15-G4: Action Icon 

– Step 3 

 

Note: If the shock cancel (Wh

en the patients’ ECG rhythm is changed to a non-shockable 

rhythm,  when  the  pads  are  disconnected  or  when  the  impedance  or  mode  is 
changed) has been occurred 3 times, the AED is performing the CPR process. 

Note:  If  “Shock  button  not  pressed.”,  the  AED  will  move  to  step  4  which  demonstrate 

CPR process directly. 

 

Step 4 

 

“It is safe to touch the patient.” 

 

“Begin  CPR.”  {Beep,  (30  times  at  a  rate  of  120/min)}  or  “If  needed,  Begin  CPR.” 

{Beep, (30 times at a rate of 120/min)} 

“Give two breaths.” {Repeat 5 times} 

 

“Stop CPR.” 

 

“Adult pads” or “Pediatric pads” 

 

Figure 22. HeartOn A15-G4: Action Icon 

– Step 4 

 

Note: If the shock cancel (

When the patients’ ECG rhythm is changed to a non-shockable 

rhythm,  when  the  pads  are  disconnected  or  when  the  impedance  or  mode  is 
changed) has been occurred 3 times, the AED is performing the CPR process. 

Note:  If  the  pads  connector  has  been  removed  in  Step  4  (CPR),  starting  from  Pad 

disconnect  icon  after  CPR.  If  not,  starting  from  Step  2  and  the  voice  prompt  for 
type of pads will be emitted. 

 

Содержание HeartOn A15

Страница 1: ...S S A Bd G n ral Wahis 53 1030 Brussels Belgium Manufacturer Mediana Co Ltd 132 Donghwagongdan ro Munmak eup Wonju si Gangwon do Korea Tel 82 2 542 3375 82 33 742 5400 Fax 82 2 542 7447 82 33 742 5483...

Страница 2: ...d by accidents or natural disasters earthquakes flooding etc Damage resulting from usage where caution statements and operating instructions shown in this manual have not been followed Damage due to n...

Страница 3: ...ed communication port 17 Setting up the AED 18 Temperature status 18 BATTERY OPERATION 21 Operating the AED on Battery Power 21 Battery Status Indication 23 Self Test 24 USING THE AED 27 AHA ERC guide...

Страница 4: ...on Icon Step 6 33 Figure 15 HeartOn A15 Action Icon Step 7 34 Figure 16 HeartOn A15 Action Icon Step 8 34 Figure 17 HeartOn A15 G4 Pad disconnect icon 35 Figure 18 HeartOn A15 G4 Action icon Step 1 36...

Страница 5: ...rical Code Articles 500 503 In accordance with the IEC EN 60601 1 Classifications the AED is not to be used in the presence of flammable substance air mixtures The AED has been designed to work on unr...

Страница 6: ...e heart Pay attention to possibility of contact with conduction part of electrode lead line cable connector other patient installation part for patient safety Do not use this AED near or within puddle...

Страница 7: ...r stacked with other equipment verify proper operation prior to use Handling or transporting the patient during ECG analysis can cause incorrect or delayed diagnosis If the AED gives a SHOCK ADVISED p...

Страница 8: ...8 This page is intentionally left blank...

Страница 9: ...oads including the pavement public squares parks subway station government buildings beaches public libraries privately owned buildings or property opened to public visible from sidewalks and any shar...

Страница 10: ...Standard 8 Action Icons HeartOn A15 G4 A15M G4 0 E AED Standard 4 Action Icons Note The alphabet E can be added as the last digit of reference number in accordance with the region Features for the AE...

Страница 11: ...DESCRIPTION OF THE AED Top and Right Panel Components Figure 1 HeartOn A15 Top and Right Panel Components Figure 2 HeartOn A15 G4 Top and Right Panel Components 1 2 3 4 5 6 7 9 10 8 2 3 4 5 6 7 1 9 10...

Страница 12: ...de between adult and pediatric infant child patient mode by pushing the patient mode switch 6 Shock button When preparation for electric shock is completed the shock button will flash Push the Shock b...

Страница 13: ...mark Battery status Environmental shipping storage atmospheric pressure limitations Temperature status Environmental shipping storage humidity limitations Contains no latex Environmental shipping sto...

Страница 14: ...14 This page is intentionally left blank...

Страница 15: ...seals and validity of use by date List of Components The following items are accessories in the package Optional accessories may be ordered if needed Contact qualified service personnel or your local...

Страница 16: ...the SD card into the SD card port 4 Close the SD card port cover 5 If necessary to update the AED through SD card turn on the AED by pushing Slide button to the right 6 After complete the update autom...

Страница 17: ...d could be used for further clinical assessment Please make sure that event data are securely saved in any of storage format when event data is accessed by the HeartOn AED Event Review Software or is...

Страница 18: ...value for environmental operation condition for long time it will be longer than usual to recognize the temperature It is recommended that AED should be stored in environmental operation condition de...

Страница 19: ...he voice prompt Unit ok and then check the AED is turned off Install 3 Place the AED into its Soft Carry Case Install 4 Put into a storage or safe visible location Note Storage differs in some countri...

Страница 20: ...20 This page is intentionally left blank...

Страница 21: ...carded batteries may explode during incineration Dispose used batteries properly Do not dispose of batteries in refuse containers Check battery capacity regularly Replace the new battery if you need E...

Страница 22: ...le battery by using the hook as shown in the first figure of the Figure 6 4 With the handle battery being connected to the AED by hook connect handle battery connector and hook of the AED as shown in...

Страница 23: ...fe time specified in above Note After 200 times of shock the voice prompt Low battery insert fresh battery will be emitted Note Due to the physical dimensions of the battery compartment only batteries...

Страница 24: ...er interaction If there is an error the status indicator displays X The self test will test your AED and ascertain if its basic functions are running Daily self test MCU and Memory RAM ROM integrity B...

Страница 25: ...icator displays X If the AED does not function properly do not use the AED Instead contact qualified service personnel or your local supplier You can also skip the battery insertion self test try foll...

Страница 26: ...26 This page is intentionally left blank...

Страница 27: ...ealth care provider only If the person is unresponsive but you are unsure that they have suffered from a SCA begin CPR When appropriate apply the AED and follow the voice prompts AHA ERC guidelines Re...

Страница 28: ...er should call for help before checking for breathing and trained rescuers are encouraged to simultaneously perform checking for pulse step After call for help Send for AED Person not responsive Not b...

Страница 29: ...ation over folds of skin such as those underneath the breast or on obese patients Excessive hair poor adhesion or air under pads may produce burns or ineffective energy transfer To apply the pads to p...

Страница 30: ...switched to right Pediatric pads If the pad is not inserted you will hear the voice prompt Plug in pads Insert connector firmly Figure 7 Patient mode switch Note The patient mode can switch even if an...

Страница 31: ...apy Note If you hear following voice prompt when the AED is turned on you should perform the following actions Low battery insert fresh battery After replace the new battery open the door case to turn...

Страница 32: ...actly as shown in the picture Press pads firmly to patient s bare chest When pads are disconnected to the AED the following voice prompt will be emitted Plug in pads Insert connector firmly When the p...

Страница 33: ...ice prompt will be emitted Shock cancelled Figure 14 HeartOn A15 Action Icon Step 6 Note If No shock advised the AED will move to step 8 which demonstrate CPR process directly Note The AED performs th...

Страница 34: ...de is changed has been occurred 3 times the AED is performing the CPR process Note If Shock button not pressed the AED will move to step 8 which demonstrate CPR process directly Step 8 It is safe to t...

Страница 35: ...en the AED is turned on you should perform the following actions Low battery insert fresh battery After replace the new battery open the door case to turn on the AED and after complete the power on se...

Страница 36: ...area where the pads are about to be applied Remove clothes from the patient s chest Place pad exactly as shown in the picture Press pads firmly to patient s bare chest Figure 18 HeartOn A15 G4 Action...

Страница 37: ...ure 20 HeartOn A15 G4 Action icon Step 2 Note If No shock advised voice prompt is emitted the AED will move to step 4 which demonstrate the CPR process Note The AED performs the Step 2 directly when i...

Страница 38: ...een occurred 3 times the AED is performing the CPR process Note If Shock button not pressed the AED will move to step 4 which demonstrate CPR process directly Step 4 It is safe to touch the patient Be...

Страница 39: ...he recommended compression rate is 100 120 beats per minute the recommended compression depth is at least 2 inches 5 cm but not more than 2 4 inches 6 cm the recommended compression ventilation ratio...

Страница 40: ...40 This page is intentionally left blank...

Страница 41: ...tteries or Li ion batteries together as they might leak electrolyte heat or explode Do not mistreat the battery or use the battery in applications not recommended by Mediana Keep the battery out of re...

Страница 42: ...Note The AED should be disposed of separately from the municipal waste stream via designated collection facilities appointed by the government or the local authorities Note The correct disposal of yo...

Страница 43: ...ay need to replace the battery with a fresh battery If the battery is not inserted into the AED the battery has a shelf life 2 years from manufacture date For diagnosis of the reason for status indica...

Страница 44: ...r replacements Figure 23 Use by date of Pads 3 Check the AED and accessories for damage or use by date Replace any accessories found to be damaged or that have exceeded their use by date 4 Check the e...

Страница 45: ...ng correctly contact qualified service personnel or your local supplier or Mediana directly for support info mediana co kr 1 There is no response to the opening the cover of the AED A CPU module may b...

Страница 46: ...ectrical appliances and high definition television it is possible that high levels of such interference due to close proximity or strength of a source may affect AED operation The AED is designed for...

Страница 47: ...gnals in the heart is often the cause of SCA but a shock can be administered to re establish sinus rhythm This treatment is called defibrillation The AED is designed to automatically detect ventricula...

Страница 48: ...victim s heart to return to a sinus rhythm HeartOn A15 The AED is a semi automatic device used for the delivery of external defibrillation therapy to resuscitate victims of SCA who are unresponsive ar...

Страница 49: ...equal to 0 2 mV V T Adult greater than or equal to 150 bpm Pediatric Infant Child greater than or equal to 180bpm Lead connection Confirm the connection and emit voice prompt Filter 0 5 to 30 Hz Indi...

Страница 50: ...IEC60529 Note The Standby condition indicates that the self test periodically runs with installed battery in the AED Storage in shipping container Temperature 20 to 60 C 4 to 140 F Relative Humidity 5...

Страница 51: ...to 109 4 F Relative Humidity 5 to 95 RH Non condensing Battery Battery Type LiMnO2 Disposable Long Life Primary Cell Voltage Capacity 15V 4200 mAh Shelf Life in the original packaging 2 years from man...

Страница 52: ...ld defibrillation Load Resistance Pulse width 1 ms Pulse width 2 ms Delivered Energy J 25 3 3 3 1 51 50 4 7 4 1 50 75 6 7 4 7 49 100 8 3 5 9 51 125 9 7 6 7 50 150 11 3 7 9 49 175 11 7 8 7 47 200 11 7...

Страница 53: ...81 82 Non shockable TN 32291 17518 1267 7062 58138 Non shockable FP 0 2 8 33 43 Specificity 100 00 99 99 99 37 99 53 99 93 Positive Predictive Value 100 00 91 67 97 24 97 99 100 00 98 55 Database for...

Страница 54: ...ty engineering to Medical devices IEC60601 1 9 2007 A1 2013 EN 60601 1 9 2008 A1 2013 Medical electrical equipment Part 1 9 General requirements for basic safety and essential performance Collateral S...

Страница 55: ...ronic instructions for use of medical devices eIFU Hazardous Substance 2015 863 EU Restriction of the use of Hazardous Substances in electrical and electronic equipment lll RoHS lll EC 1907 2006 Regis...

Страница 56: ...energy in order to perform its intended function Nearby electronic equipment may be affected RF emissions CISPR 11 Class B The AED is suitable for use in all establishments Table 6 Electromagnetic Im...

Страница 57: ...e level in each frequency range b Interference may occur in the vicinity of equipment marked with the following symbol Note At 80 MHz and 800 MHz the higher frequency range applies Note These guidelin...

Страница 58: ...wer of Transmitter in watt Separation distance according to frequency of transmitter in meter 80 MHz to 800 MHz d 1 2 P 800 MHz to 2 5GHz d 2 3 P 0 01 0 12 0 23 0 1 0 38 0 73 1 1 2 2 3 10 3 8 7 3 100...

Страница 59: ...Band 7 Pulse Modulation 217 Hz 2 1 28 5240 5500 5785 5100 5800 WLAN 802 11 a n Pulse Modulation 217 Hz 0 2 1 9 Note If necessary to achieve the IMMUNITY TEST LEVEL the distance between the transmitti...

Отзывы: