30
1.
Check the status indicator displays “O”.
2. To open the cover, push the
Slide button
to the right.
3. Turn on the AED by opening the cover.
4. The AED automatically starts the Power-On-Self Test.
5. The test result is displayed on the status indicator and the voice prompt sounds.
Self test is passed : Voice prompt
“Unit ok”, Status indicator “O”
Self test is failed : Voice prompt
“Unit failed”, Status indicator ”X”
6. If the pads are inserted and the Patient mode switch is selected, you will hear the voice
prompt:
Patient mode switch is switched to left,
“Adult pads”
Patient mode switch is switched to right,
“Pediatric pads”
If the pad is not inserted, you will hear the voice prompt:
“Plug in pads. Insert connector firmly.”
Figure 7. Patient mode switch
Note: The patient mode can switch even if any step except for CPR is going on
progress. .If the patient mode is changed, the AED will emit the voice prompt
“Adult pads” or “Pediatric pads”. When the patient mode is changed during CPR,
the AED will not emit the voice prompt.
Note: Return to ECG analysis, while analyzing ECG or delivering electric shock, even
though the patient mode is changed during process. When the number of mode
change is 3 times, the AED will proceed the CPR.
7. Verify the AED up to
‘6.’ which is activated normally and follow voice prompt and action
icon. The red LED will flash under the relevant action icon.