Defect
Source of malfunction
Corrective actions
Vacuum gauge does not fit An attempt has been made
to connect the vacuum
gauge to the tapping unit
with the scale pointing
forwards
Always insert vacuum gauge with a
90° twist and then turn it to the front
Vacuum gauge for the
incorrect pressure range
Select vacuum gauge for the correct
pressure range (see type plate);
connections are coded
Oversuction of product
despite protective system
Tapping unit and overflow
protection device mounted
at an angle
Operate tapping unit in a vertical
position only
Overflow protection device
contaminated
Clean overflow protection device
No foam inhibitor used
Use standard commercial foam
inhibitor
Unit is not suspended high
enough above the patient's
chest
Suspend unit higher or use overflow
protection device
Gas probe does not fit into
the terminal unit
Terminal unit for the wrong
gas type
Check gas type and use another
terminal unit
Tab. 5:
Corrective actions
7.4
Repairs
The following may require repairs by the manufacturer or an authorised service partner:
• Liquid has penetrated the device.
• Performance has significantly decreased.
• Inexplicable notifications appear.
• Abnormal noises occur.
• Functional faults cannot be rectified according to the measures in chapter Malfunctions and
If defects are detected, the product may not be used any longer.
Make a note of the defects and the REF number on the type plate and inform your ATMOS
representative.
Observe the information in chapter Sending in the device [
7.5
Service hotline
+49 7653 689-0
7.6
Spare parts
5752 2103
Vacuum gauge for drainage
Tab. 6:
Spare parts