Introduction
Basic requirements
GA 5752 2135 GB 18
11
1
Contraindications:
The FINA VAC D 150 may not be used for the following purposes:
• Outside the medical sector
• In MR areas
• In the home care sector
• Being operated directly by the patient
• For vacuum extraction
• For the aspiration of flammable or explosive liquids
• For the aspiration of smoke that is generated during HF and
laser surgery without the connection of an intermediate smoke
filter
• For thoracic drainages
• For surgical aspiration
• For bronchial aspiration
• With central gas supply systems with supply pressures other
than −100 kPa bis −60 kPa
The product is:
Active
Sterility:
Not a sterile product
Single-use product /
reprocessing:
The device and parts of the accessories are reusable. For
information on reprocessing, cleaning and disinfection, please see
the operating instructions.
1.6.4
Versions of the FINA VAC D 150
The connection of the FINA VAC D 150 to the terminal unit for vacuum depends on the model
being used:
Version A: Tapping unit with integrated gas pin
• FINA VAC D 150 is fitted directly to the terminal unit.
Version B: Tapping unit with rail clamp and NIST connection
• FINA VAC D 150 is designed for mounting onto a 25 x 10 mm equipment rail and is supplied
via a NIST connection with vacuum from a terminal unit connected using a connection tube
with gas probe.
Products and accessories are only permitted with ISO colour coding. In Germany, Austria and
Switzerland, products with neutral colour coding are also permitted.
NOTE
The products are supplied with ISO coding. The scope of delivery includes a label
for neutral colour coding.
The product is available in the following versions:
• FINA VAC D 150 Wall DIN ISO (REF 5752 3739)
• FINA VAC D 150 Equipment rail ISO (REF 5752 3740)