Safety notes
Product safety notes
GA 5752 2135 GB 18
15
2
WARNING!
Backflow of aspirated secretion!
In the event of oversuction, the aspirated secretion may flow back to the patient if
there is secretion still left in the suction tube.
Before replacing the septic fluid jar in the event of oversuction or switching off the
vacuum, always remove the tube from the patient first.
WARNING!
Impacts!
Impacts may cause damage to sensitive, precision mechanical components.
Do not expose the product to impacts.
WARNING!
Foaming!
Foam may be created when extracting secretion. Foam is detrimental to the
functioning of the mechanical overflow protection. This raises the risk that secretion
may penetrate the product and cause it to break down.
Use an ordinary foam inhibitor.
WARNING!
Risk to patient!
Use the special tapping unit for thoracic draining for vacuum limiting in the
regulating range less than 3 kPa, in particular in thoracic drainage.
WARNING!
Non-permissible load!
If the permissible load is exceeded, leakages may occur at the connection between
the terminal unit and the gas probe.
In accordance with DIN EN ISO 9170-1, the overall weight of the product and
accessories may not exceed 2 kg.
WARNING!
Risk of injury!
Replace the hydrophobic bacterial and viral filter immediately if it is discoloured,
contaminated or oversucked.
Furthermore, the filter must be changed if the vacuum displayed is −0.13 bar /
−13 kPa when the vacuum controller is in the ‘max’ position and the suction tube is
open.