![MEDAP 57505128 Скачать руководство пользователя страница 8](http://html1.mh-extra.com/html/medap/57505128/57505128_operating-instructions-manual_1759332008.webp)
8
Starting the operation
Before you connect the aspirator, check if the supply
voltage corresponds to the information on the type label.
• Place the aspirator in a vertical and safe position.
• Place the septic fluid jar cap (5) onto the septic fluid
jar (9).
• Use both your balls of the thumb to press down the
carrying handle and have the hooks lock into the
grooves of the septic fluid jar.
• Insert the suction unit (10) into the holder with the
patient's side connection (white) directing to the front
and connect tubing with the finger tip or the suction
element.
Only use original tubes or vacuum-tested tubes in accor-
dance with the list of accessories and pay attention to
correct tubing.
• Connect suction unit with the shorter tube via the hu-
midity filter to the aspirator (black tube nipple). Pay
attention to direction of flow! The side with the text
must point in the direction of the suction unit.
• Plug in power plug.
• Carry out a functional check. Switch on the device,
keep the finger tip closed and check if vacuum is built
up. If no or hardly any vacuum is built up, please refer
to the troubleshooting table on page 12.
Preparation of accessories
Assembling manometer with overflow prevention device
(Kit REF 57521235)
In order to regulate the vacuum and to protect the aspira-
tor from excess aspiration, the kit " manometer with
overflow prevention device" may be installed.
• Install the manometer with overflow prevention device
in accordance with the instructions ET 0659 01 / REF
57520687 (included in the kit).
Operation
!
Installation between aspirator and sep-
tic fluid container
Side with text pointing towards suction
unit!
Overflow pre-
vention device
Float