Pull the power plug prior to each cleaning. Comply with
the hygiene regulations and protect you eyes and skin
when working with disinfectants.

Do not clean the surfaces of the housing with solvents
that contain hydrocarbon (e.g. benzene, diluent).

Efficiency of disinfection and/or sterilization is
likely to be impaired by trapped or encapsula-
ted impurities or residual humidity.

• Clean parts thoroughly, rinse with adequate amount of

water an dry parts until they absolutely dry.

• Pull power plug.

• Disconnect the suction element (10) from the tube

connections and take it out of the holder.

Make sure that no residues of the septic fluid pour out of
the connecting tube.

• Remove the septic fluid jar cap (5) and empty the sep-

tic fluid jar (9) in accordance with the operator
guidelines.

Do not treat septic fluid containers made of polysulphone
with strong acids or bases. Risk of fractures due to tension.

• Always disassembly the overflow prevention device

(12) for preparation. Unscrew it from the manometer,
pull off the float cage  and remove the float.

Make sure that the sealing rims of the float are not dama-
ged.

!

Cleaning, Disinfection, Sterilization

!

!

)

1

Basic aspirator

X

X

X

9

Septic fluid jar 0.7 l

X

X

X

X

X

5

Septic fluid jar cap

16

Overflow prevention device (acc.)

6

Tube  nipple

X

X

X

X

X

X

7

O-ring

2

Silicone tube

8

Seal  EPDM

4

Finger  tip

Sterilization***

in solution**

Wiping,

spraying with disinfectant

Disinfection

Aseptor

****Attention! Sterilization with superheated steam accelerates the natural ageing of plastics.

Steam 134°C****

* Item numbers based on diagram on page 3
** Pay attention to thorough rinsing with water and subsequent drying.
*** Carefully remove contamination and residues prior to sterilization and thoroughly dry off.

Disposables

Gas

Steam 120°C****

Item*

Description

12

Содержание 57505128

Страница 1: ...1 Septic Fluid Aspirator P 7050 P 7050 100 01 57505128 REF 57520525 Valid as of August 03 Operating Instructions Sekretsauger P 7050...

Страница 2: ...sories 7 Operation 8 Starting the operation 8 Preparation of accessories 8 Aspiration 9 Cleaning disinfection sterilization 10 Troubleshooting 12 Maintenance 12 Accessories Consumables Spare Parts 13...

Страница 3: ...atients and operating staff as well as for the prevention of damage to the device and or functional defects The meaning of the pictorial signs is as follows STOP Danger for patient and operating staff...

Страница 4: ...ple 7 O Ring 8 Gasket 9 Septic fluid jar 10 Suction unit 1 3 2 4 5 6 7 8 10 9 Humidity filter Sekretsauger P 7050 Accessories Manometer with overflow prevention device 11 Manometer 12 Overflow prevent...

Страница 5: ...tion according to instruction by qualified personnel If other than original parts are used performance and safety might be impaired The device must be used in vertical position only Do not switch on t...

Страница 6: ...e P 7050 aspirator the vacuum is generated by an electrically driven pumping aggregate The vacuum is 0 8 bar and can be regulated with a ma nometer with regulating valve and overflow prevention device...

Страница 7: ...contai ner the tube nipple for the connection to the vacuum source must fit for the connecting tube Suction tube Vacuum resistant to 0 8 bar must not collapse inner diameter of tube must correspond t...

Страница 8: ...via the hu midity filter to the aspirator black tube nipple Pay attention to direction of flow The side with the text must point in the direction of the suction unit Plug in power plug Carry out a fu...

Страница 9: ...be Adjust vacuum Close the finger tip control 4 with a finger for continuous operation close suction opening with the plug Aspirate septic fluid If frothing occurs during aspiration use standard froth...

Страница 10: ...ot treat septic fluid containers made of polysulphone with strong acids or bases Risk of fractures due to tension Always disassembly the overflow prevention device 12 for preparation Unscrew it from t...

Страница 11: ...e are only valid for original Heraeus Med components If you use components of other manufacturers follow the procedures of the respec tive manufacturer Make sure that no fluid penetrates the aspirator...

Страница 12: ...eck fuses Power plug does not correctly fit Insert power plug correctly Power cable defective Motor does not run Septic fluid jar cap not correctly placed Place septic fluid jar cap correctly Humidity...

Страница 13: ...630 Spare parts This list contains spare parts which can be replaced by the final user A com plete spare part list will be supplied to especially trained and qualified personnel Septic fluid jar 0 7 l...

Страница 14: ...on Transportation storage conditions Temperature 15 C to 50 C Rel air humidity 100 rel humidity Operating conditions Temperature 15 C to 30 C Air humidity 30 to75 rel humidity Performance data Art no...

Страница 15: ...cture of P 7050 materials are used which are utilizable or can be recycled and which do not carry a special or extraordinary risk during disposal Heraeus Med will recycle a device which has become un...

Страница 16: ...Rastatt D 76437 Rastatt AUSGABE AUGUST 2003 SECTION AUGUST 2003 REF 5752 0525 REF 5752 0525 VERTRIEB 0 18 03 21 21 33 SALES 49 72 22 932 644 VERTRIEB FAX 0 18 03 21 21 77 FAX 49 72 22 932 571 KUNDEND...

Отзывы: