6
Correct operation
The Septic Fluid Aspirator P 7050 is a small aspirator
which is suitable for low flow rates and high vacuum,
also in continuous operation. It is used for the aspiration
of body fluids (septic fluids, blood and serous fluids) and
particles contained therein from natural and artificial ori-
fices and is intended for the temporary application on the
patient in the case of septic fluid aspiration. The aspirator
was designed for use in hospitals, medical practices and
nursing homes and is to be operated by trained and quali-
fied staff. It can also be used in home care after the user
has been adequately trained and instructed.
Contra-indications:
The septic fluid aspirator P 7050 is
neither
suitable for
use in surgery nor as a drainage aspirator. The particles
contained in the gas phase might lead to an early clogging
of the humidity filter.
Technical description
In the P 7050 aspirator, the vacuum is generated by an
electrically driven pumping aggregate.
The vacuum is - 0,8 bar and can be regulated with a ma-
nometer with regulating valve and overflow prevention
device (accessory).
The overflow prevention device protects the aggregate
from excess aspiration. If the filling level is high, the float
rises in the overflow prevention device, closes off the
suction tube and thus prevents fluids to flow out of the
septic fluid container.
Please check the delivered goods after unpacking for
completeness and compliance with the order. Also refer
to your order for the supply schedule. The standard supp-
ly schedule includes the basic aspirator with power cable
and a 0.7l suction unit without overflow prevention de-
vice (available as accessory) as well as the tubing incl.
humidity filter and a finger tip. You can separately order a
manometer with regulating valve (regulating unit). If you
have any complaints please contact your supplier.
Prior to the first use, the device has to be prepared in ac-
cordance with the hygiene guidelines. (
Ö
Cleaning, ...).
General Description