These operating instructions are intended for:
operating staff
the user at home.

The operating instructions do not replace the careful training
of personnel or user.

Only qualified personnel are permitted to use
this device in accordance with the operating
instructions.

During aspiration of septic fluids, the applicable
hygiene regulations must be complied with.

Danger of fire

The aspirator must not be used in AP-areas where there is
a danger of explosions (e.g. in the vicinity of anaesthetic
equipment).

Power supply

The system voltage must comply with the information on the
type label.
Pull the power plug before you open or clean the device.
Fluids must never penetrate the parts in circuit.

Clinical application

according to instruction by qualified personnel.

If other than original parts are used, performance and safety
might be impaired.

The device must be used in vertical position only.

Do not switch on the aspirator when vacuum is present
since the device might overheat.

Make sure that the aspirator has a safe stand.

In case of damages (e.g. falls) have the aspirator checked for
its operational safety and have it repaired if required.

The manufacturer does not assume any liability for other ap-
plications than those specified in the user instructions or for
modifications on your own initiative. Repairs must exclusive-
ly be carried out by the manufacturer or authorized qualified
persons, using original components and complying with the
valid safety regulations for electrical installation.

All repair work for which the use of tools is required must be
carried out by the technical service.

We care about safety!

!

Содержание 57505128

Страница 1: ...1 Septic Fluid Aspirator P 7050 P 7050 100 01 57505128 REF 57520525 Valid as of August 03 Operating Instructions Sekretsauger P 7050...

Страница 2: ...sories 7 Operation 8 Starting the operation 8 Preparation of accessories 8 Aspiration 9 Cleaning disinfection sterilization 10 Troubleshooting 12 Maintenance 12 Accessories Consumables Spare Parts 13...

Страница 3: ...atients and operating staff as well as for the prevention of damage to the device and or functional defects The meaning of the pictorial signs is as follows STOP Danger for patient and operating staff...

Страница 4: ...ple 7 O Ring 8 Gasket 9 Septic fluid jar 10 Suction unit 1 3 2 4 5 6 7 8 10 9 Humidity filter Sekretsauger P 7050 Accessories Manometer with overflow prevention device 11 Manometer 12 Overflow prevent...

Страница 5: ...tion according to instruction by qualified personnel If other than original parts are used performance and safety might be impaired The device must be used in vertical position only Do not switch on t...

Страница 6: ...e P 7050 aspirator the vacuum is generated by an electrically driven pumping aggregate The vacuum is 0 8 bar and can be regulated with a ma nometer with regulating valve and overflow prevention device...

Страница 7: ...contai ner the tube nipple for the connection to the vacuum source must fit for the connecting tube Suction tube Vacuum resistant to 0 8 bar must not collapse inner diameter of tube must correspond t...

Страница 8: ...via the hu midity filter to the aspirator black tube nipple Pay attention to direction of flow The side with the text must point in the direction of the suction unit Plug in power plug Carry out a fu...

Страница 9: ...be Adjust vacuum Close the finger tip control 4 with a finger for continuous operation close suction opening with the plug Aspirate septic fluid If frothing occurs during aspiration use standard froth...

Страница 10: ...ot treat septic fluid containers made of polysulphone with strong acids or bases Risk of fractures due to tension Always disassembly the overflow prevention device 12 for preparation Unscrew it from t...

Страница 11: ...e are only valid for original Heraeus Med components If you use components of other manufacturers follow the procedures of the respec tive manufacturer Make sure that no fluid penetrates the aspirator...

Страница 12: ...eck fuses Power plug does not correctly fit Insert power plug correctly Power cable defective Motor does not run Septic fluid jar cap not correctly placed Place septic fluid jar cap correctly Humidity...

Страница 13: ...630 Spare parts This list contains spare parts which can be replaced by the final user A com plete spare part list will be supplied to especially trained and qualified personnel Septic fluid jar 0 7 l...

Страница 14: ...on Transportation storage conditions Temperature 15 C to 50 C Rel air humidity 100 rel humidity Operating conditions Temperature 15 C to 30 C Air humidity 30 to75 rel humidity Performance data Art no...

Страница 15: ...cture of P 7050 materials are used which are utilizable or can be recycled and which do not carry a special or extraordinary risk during disposal Heraeus Med will recycle a device which has become un...

Страница 16: ...Rastatt D 76437 Rastatt AUSGABE AUGUST 2003 SECTION AUGUST 2003 REF 5752 0525 REF 5752 0525 VERTRIEB 0 18 03 21 21 33 SALES 49 72 22 932 644 VERTRIEB FAX 0 18 03 21 21 77 FAX 49 72 22 932 571 KUNDEND...

Отзывы: