Med-link Electronics COX801VB Скачать руководство пользователя страница 8

11

8.4.1 SpO2

    Blood oxygen saturation (SpO2) refers to the percentage of blood oxygen-bound hemoglobin 
(HbO2) in the total binding hemoglobin (Hb, hemoglobin concentration), which is an important 
physiological parameter of respiratory circulation;
    Spo2 alarming high limit value "---" means no ultra-high limit alarming value; SpO2 alarming 
low limit value "90" means ultra- low limit alarming value;
    If the measured Spo2 value is higher than the high limit value, the measured value turns 
yellow, triggering an medium-level alarm;
    If the measured Spo2 value is lower than the low limit value, the measured value turns red, 
triggering an high-level alarm.

8.4.2 PR

    Shows the number of beats of pulse per minute under normal measurement;
    PR alarming high limit value "150" indicates the ultra-high limit alarming value; PR alarming 
low limit value "50" indicates the ultra-low limit alarming value;
    If the measured pulse rate value is higher than the high limit value, the measured value turns 
red, triggering an high-level alarm;
    If the measured pulse rate value is lower than the low limit value, the measured value turns 
red, triggering an high-level alarm.

8.4.3 PI

    Shows current blood flow perfusion capacity under normal measurement conditions.

8.4.4 Bar graph display area

    Pulse amplitude indication under normal measurement.

8.4.5 Plethysmographic display area 

    Demonstrate pulse fluctuations under normal measurement. 

8.4.6 Module Status Display Area

     Display animal information under normal conditions, such as: ID: 01 gender Male Age 28;
    Display the status of the module under abnormal measurement, such as: "The probe is not 
connected", "The probe fault", etc.

8.4.7 Date display area

    Display the current date of the device.

8.4.8 Time display area

    Display the current time of the device.

8.4.9 Battery display area

    Display the remaining battery power, there is a low battery prompt;
    If it is less than 3.65V, the battery symbol will turn red to indicate low battery;
    If it is less than 3.3V, it will automatically shut down.

8.4.10 Blood oxygen pulse rate trend graph display area

    Choose to show trends in blood oxygen and pulse rate for different periods of time.

Note: 

(1) The normal update period of the data is 1 second, and the invalid value is displayed if the data 
is not updated for more than 30 seconds. 
(2) The influence of other signal processing on the display and transmission of parameters is 
within 1 second. 

8.4 Explanation of measurement results

Figure 12

12

9   Alarming system

(1) alarming signal generated by alarming system is divided into visual alarming signal and 
auditory alarming signal. 
(2) the visual alarming signal has indicator light alarming indication and display text alarming 
indication. 
(3) the alarming system is divided into three levels: high-level, medium-level, and low-level 
alarms, and the alarming level cannot be changed. 
(4) in addition, the alarming system also provides prompt information. 
(5) the maximum delay time of the alarming system is less than 2 seconds. 

9.1 Overview

There are different between using a pulse carbon oximeter sensor and pulse oximeter sensor will 
be different.
After entering the setup interface,
Audible alarm: The alarming sound is not broadcast after the alarming sound is broadcast.
The text displaying alarm: it is not displayed because it enters the setting interface.
Indicator alarm:  always prompt the latest, highest priority alarm before entering the setup 
interface and no longer update.

9.2 Precautions

● The improper use of Probe;
● An  important  indicator  of  hemoglobin  dysfunction  (such  as  carbonaceous  hemoglobin  or 

methemoglobin);

● Intravascular staining exceeding the standard, such as indole cyan green or methylene blue;
●  Strong  ambient  light  such  as  surgical  lights  (especially  Xenon  lamps),  bilirubin  lamps, 

fluorescent lamps, infrared light; 

● High frequency electrosurgical equipment and defibrillators;
● Rhythmic beating of vein;
● Tight Spo2 sensor; 
● Low perfusion index.

8.5 The inaccurate measurement results may be due to 

please use the attachment recommended in this manual(see attached list). 

8.6 Warning

After the measurement is over, turn off the instrument. 

8.7 End of measurement

Appendix alarming indicator 

characteristics; Appendix 

alarming sound characteristics;

Display text alarming indication

"Probe is not connected".

The screen 

display 

"The probe is 

not connected"

Probe is not 

connected

Appendix alarming indicator 

characteristics; Appendix 

alarming sound characteristics;

Display text alarming indication

“Probe current is too large".

The screen 

display

"The probe is 

current is too 

large"

Probe current 

is too large

Appendix alarming indicator 

characteristics; Appendix 

alarming sound characteristics;

Display text alarming indication

"Probe fault".

The screen 

display 

"The probe 

fault"

Probe fault

High-level 

alarm

Technical 

alarm

See the following table:

Host uses 

Spo2 

sensor

Содержание COX801VB

Страница 1: ...eter COX801VB manual for user Please read this manual carefully before using Shenzhen Med link Electronics Tech Co Ltd Address 4th and 5th Floor Building Two Hualian Industrial Zone Xinshi Community D...

Страница 2: ...ent process 8 3 Start measuring 8 4 Interpretation of measurement results 8 4 1 SpO2 8 4 2 PR 8 4 3 PI 8 4 4 Bar chart display area 8 4 5 Trace wave display area 8 4 6 Module status display area 8 4 7...

Страница 3: ...fied by our company see Appendix P Please note that this product is only used for veterinary It can not be used for human clinical If it is used for human rather than animal by the users and caused th...

Страница 4: ...ld for 2 seconds to turn off the power when the power is turned on Figure 1 Type C 4 3 general description of the parameters 4 Working principle COX801VB can continuously measure arterial oxygen satur...

Страница 5: ...down key to adjust the value of year then press the set key to switch month day hour minute successively for setting The time shows the 24 hour system 6 1 Time setting 1 Under the setting interface pr...

Страница 6: ...t alarming value of PR 1 If the upper limit of the alarming line is not valid i e the lower limit of the alarming line can be arbitrarily selected from 30 bpm to 350 bpm or invalid i e 2 If the upper...

Страница 7: ...if you are allergic to silicone and ABS plastics Please dispose of used mainframes accessories and packaging materials in accordance with local laws and regulations Figure 7 Figure 8 Correct connecti...

Страница 8: ...her signal processing on the display and transmission of parameters is within 1 second 8 4 Explanation of measurement results Figure 12 12 9 Alarming system 1 alarming signal generated by alarming sys...

Страница 9: ...m level alarm Yellow low level alarm Green Red Green 2 Attention the alarming level is not changed Statement the display text alarm indicates that it can be clearly observed at a distance of 1m Physio...

Страница 10: ...2 lower limit of alarm PR Upper limit of alarm PR lower limit of alarm Battery is too low no 90 140BPM 50BPM Appeared below about 3 65V Long term transportation storage over 3 months 15 10 specificati...

Страница 11: ...hrough TYPEC cable Battery is too low No alarm Alarm pause alarm shutdown or horn failure Please check if the alarm tone is off Please check if the set volume is level 0 Invalid button Internal cause...

Страница 12: ...tions or use concentration as low as possible 2 Do not submerge the device in liquid 3 Do not pour liquid onto equipment or accessories 4 Do not allow liquid to enter the casing 5 Do not use abrasive...

Страница 13: ...warranty period of our equipment is set according to the electronic product after sales service standards stipulated by national laws The warranty period is one year for mainframe and three months fo...

Страница 14: ...nly valid in China one card no replacement 7 The accuracy of this product is medical standard To ensure the accuracy of your use after the warranty period the company promises to continue to provide l...

Отзывы: