Warning

7

Overview:

The alarming system of this product has only a set of alarm preset alarm limits. See the factory 
default alarming limit (Appendix H) for details.
Alarming limits can be changed and saved with the following alarming limit settings.
Alarming limit value previously saved will be used by default for each start-up of the power-on 
alarming system.
If you want to restore the factory default alarming limit, please refer to <<Reset to Factory 
Settings>> of <<System Settings>>.

(1)Under the main menu interface, press the set/confirm button for 2 seconds to enter the 
alarming Settings menu, as shown in the following figure;
(2)The short press the set/confirm button can switch the alarming value to be set, and the order 
is as follows: 
SpO2 upper limit alarming value-> SpO2 lower limit alarming value-> PR upper limit alarming 
value-> PR lower limit alarming value-> Exit and save alarming value Long Press set/confirm 
button can exit and save the alarming value setting, press the up/down button can adjust the 
corresponding alarming value.

7  

The host uses Spo2 sensor alarming settings

(1) If the lower limit of the alarming line is not valid (i.e., "--"), the upper limit of the alarming line 
can be arbitrarily selected from 40% to 99% or invalid (i.e., "--").
(2) If the lower limit of alarming line is 40% ~ 97%, the upper limit of alarming line can be 
selected from (lower limit of alarming line +2) ~ 99% or invalid (i.e., "--"). 
(3) If the lower limit of alarming line is 98% ~ 99%, the upper limit of alarming line can only be 
invalid (i.e., "--").

7.1 Adjustment range of SpO2 upper limit alarming value 

(1) If the upper limit of the alarming line is not valid (i.e., "--"), the lower limit of the alarming line 
can be arbitrarily selected from 40% to 99% or invalid (i.e., "--").
(2) If the upper limit of alarming line is 42% ~ 99%, the lower limit of alarming line can be 
selected from 40% ~ (upper limit of alarm line-2) or invalid (i.e., "--").
(3) If the upper limit of the alarming line is between 40% and 41%, the lower limit of the alarming 
line can only be invalid (that is, the "--").

7.2 Adjustment range of SpO2 lower limit alarming value

Warning: 

Please observe whether the current alarming limit is in compliance with 

                this animal before use.

Note: 

When the power supply loses at any time, powering up again and starting  

          the alarming system default to using the previously saved alarming limits.

Figure 6

Warning

8

8   Measurement

1) If the lower limit of the alarming line is not valid (i.e., "--"), the upper limit of the alarming line 
can be arbitrarily selected from 30 bpm to 350 bpm or invalid (i.e., "--").
2) If the lower limit of the alarming line is 30 bpm ~ 348 bpm, the upper limit of the alarming line 
can be selected from (lower limit of alarm line +2) ~ 350 bpm or invalid (i.e., "--"). 
3) If the lower limit of the alarming line is 349 bpm ~ 350 bpm, the upper limit of the alarming line 
can only be invalid (i.e., the "--"). 

Statement: 

Please set the system and alarming limit before starting the measurement.

7.3 Adjustment range of upper limit alarming value of PR 

1) If the upper limit of the alarming line is not valid (i.e., "--"), the lower limit of the alarming line can 
be arbitrarily selected from 30 bpm to 350 bpm or invalid (i.e., "--"). 
2) If the upper limit of the alarming line is 32 bpm ~ 350 bpm, the lower limit of the alarming line 
can be selected from 30 bpm ~ (upper limit of alarm line-2) or invalid (i.e., "--"). 
3) If the upper limit of the alarming line is 30 bpm to 31 bpm, the lower limit of the alarming line 
can only be invalid (i.e., the "--"). 

7.4 Adjustment range of lower limit alarming value of PR

(1)The sensor used in this equipment has a wavelength of red light: 660 nm, infrared light: 905 nm. These 
light waves cannot directly illuminate the animal's eyes to avoid injury. When a  treatment method similar 
to photodynamic therapy is used during the measurement, the wavelength may be interfered to affect the 
measurement.
(2)Periodically inspect the main components and accessories of the oximeter to confirm that there are no 
obvious damages that may affect the safety and monitoring performance of animal. It is recommended to 
check the product before each use. If the oximeter has obvious damage, stop using the product.
(3)when monitoring the tested animals for a long time, the position of the Spo2 sensor attached should be 
checked every 2 hours, and the proper movement should be carried out when the skin changes. The 
same measuring point should be measured continuously for up to 4 hours. Some tested animals may 
need to be examined more frequently, such as those with perfusion disorders or skin sensitivity. Because 
prolonged monitoring may increase unpredictable skin changes, such as allergies, redness, blistering or 
compression necrosis. 
(4)If there is an emergency to turn off the equipment, remove the probe from the sensor to stop the test, 
unplug all the accessories in the equipmen and turn off.
(5)Please carefully check the equipment before using this product to confirm that it can be used normally.
(6)The sensor of this oximeter should not be used in the probe or other parts of the arterial catheter, blood 
pressure cuff or intravenous injection.
(7)Make sure that the light source comes from the interior of the transparent silica gel. In addition, the light 
source will affect the measurement results, including fluorescent lamps, red light source ,infrared water 
heaters, direct sunlight and so on.
(8)Frequent activity of the subject or interference from external radio can affect the accuracy of the test.
(9)Only use the sensor specified in this manual and follow the instructions for use. Observe all warnings 
and precautions.
(10)High frequency electrosurgical equipment and defibrillators can affect the use of the equipment.
(11)Use the equipment facing the operator and within 1 meter to view the measurement parameters.

8.1 Precautions

Warning:

If different alarming line values are used for the same or similar equipment in any single 
area, there may be dangerous. For example, intensive care unit,  heart operating room 
and so on. Please set different alarming limits according to the condition of each tested 
animal, and ensure its safety and effectiveness.

Содержание COX801VB

Страница 1: ...eter COX801VB manual for user Please read this manual carefully before using Shenzhen Med link Electronics Tech Co Ltd Address 4th and 5th Floor Building Two Hualian Industrial Zone Xinshi Community D...

Страница 2: ...ent process 8 3 Start measuring 8 4 Interpretation of measurement results 8 4 1 SpO2 8 4 2 PR 8 4 3 PI 8 4 4 Bar chart display area 8 4 5 Trace wave display area 8 4 6 Module status display area 8 4 7...

Страница 3: ...fied by our company see Appendix P Please note that this product is only used for veterinary It can not be used for human clinical If it is used for human rather than animal by the users and caused th...

Страница 4: ...ld for 2 seconds to turn off the power when the power is turned on Figure 1 Type C 4 3 general description of the parameters 4 Working principle COX801VB can continuously measure arterial oxygen satur...

Страница 5: ...down key to adjust the value of year then press the set key to switch month day hour minute successively for setting The time shows the 24 hour system 6 1 Time setting 1 Under the setting interface pr...

Страница 6: ...t alarming value of PR 1 If the upper limit of the alarming line is not valid i e the lower limit of the alarming line can be arbitrarily selected from 30 bpm to 350 bpm or invalid i e 2 If the upper...

Страница 7: ...if you are allergic to silicone and ABS plastics Please dispose of used mainframes accessories and packaging materials in accordance with local laws and regulations Figure 7 Figure 8 Correct connecti...

Страница 8: ...her signal processing on the display and transmission of parameters is within 1 second 8 4 Explanation of measurement results Figure 12 12 9 Alarming system 1 alarming signal generated by alarming sys...

Страница 9: ...m level alarm Yellow low level alarm Green Red Green 2 Attention the alarming level is not changed Statement the display text alarm indicates that it can be clearly observed at a distance of 1m Physio...

Страница 10: ...2 lower limit of alarm PR Upper limit of alarm PR lower limit of alarm Battery is too low no 90 140BPM 50BPM Appeared below about 3 65V Long term transportation storage over 3 months 15 10 specificati...

Страница 11: ...hrough TYPEC cable Battery is too low No alarm Alarm pause alarm shutdown or horn failure Please check if the alarm tone is off Please check if the set volume is level 0 Invalid button Internal cause...

Страница 12: ...tions or use concentration as low as possible 2 Do not submerge the device in liquid 3 Do not pour liquid onto equipment or accessories 4 Do not allow liquid to enter the casing 5 Do not use abrasive...

Страница 13: ...warranty period of our equipment is set according to the electronic product after sales service standards stipulated by national laws The warranty period is one year for mainframe and three months fo...

Страница 14: ...nly valid in China one card no replacement 7 The accuracy of this product is medical standard To ensure the accuracy of your use after the warranty period the company promises to continue to provide l...

Отзывы: