MED Associates PHM-104A Скачать руководство пользователя страница 19

MED ASSOCIATES, INC. 

PHM-104A HIGH/LOW SPEED PRECISION PUMP 

 

28-V Output Via MED-PC 

This section describes the procedure for using a MED-PC 28-Volt output to turn the pump on and off.  
The pump will operate using the default parameters.   

1.

 

Using the 24 VDC power supply connect the 

24 VOLTS

 connector on the PHM-104A to a standard 

wall outlet. 

2.

 

Turn the PHM-104A on using the green ON/OFF switch. 

3.

 

Set the pump’s default parameters using the pump control application described in  Figure  3.4  - 
PHM-104A Control Software Screen.
 

4.

 

Connect the 

MED OPERATE

 port (Figure 2.3) on the pump to any available output on a standard 

MED Connection Panel (e.g. SG-716D), using the included cable. 

NOTE:  Be sure that the USB cable is NOT connected to the pump, as the MED OPERATE port will not 
be recognized if it is. 

5.

 

Set the FLUSH/REMOTE/TEST switch to 

REMOTE

.  The pump will now execute a complete infusion 

using the default parameters each time it receives a 28-Volt output pulse from MED-PC. 

 

 

 

- 15 -

 

DOC-187 Rev. 1.6 Copyright © 2019 

Med Associates, Inc. 

Содержание PHM-104A

Страница 1: ...or research HIGH LOW SPEED PRECISION PUMP PHM 104A USER S MANUAL DOC 187 Rev 1 6 Copyright 2019 All Rights Reserved Med Associates Inc P O Box 319 St Albans Vermont 05478 Phone 802 527 2343 Fax 802 52...

Страница 2: ...This page intentionally left blank...

Страница 3: ...ose Capabilities 4 Operating Instructions 6 Chapter 3 Syringe Placement 6 Hardware Setup 7 Software Controlled Operation 8 MED PC Controlled Operation 10 USB Control Via MED PC 10 Sample MED PC Progra...

Страница 4: ...______________ _____________________________________________________________________________________ _____________________________________________________________________________________ _____________...

Страница 5: ...quality Hamilton 1700 Series gastight glass syringes to ensure precise accurate delivery The pump supports three syringe sizes 10 l 100 l and 500 l A wide range of injection rates is possible as well...

Страница 6: ...position Enables the MED OPERATE connector When the switch is in TEST position Completes one infusion at the default pump settings refer to Table 3 1 PHM 104A Control Software User Interface Set Defau...

Страница 7: ...IATES INC PHM 104A HIGH LOW SPEED PRECISION PUMP Figure 2 2 Manifold End of the PHM 104A Figure 2 3 Control End of the PHM 104A Figure 2 4 Syringe Components 3 DOC 187 Rev 1 6 Copyright 2019 Med Assoc...

Страница 8: ...on rate included in the highlighted portion of each graph is attainable Using Figure 5 10 uL syringe as an example a dose of 3 uL may be infused at a rate up to 27 uL sec Dose capabilities are based o...

Страница 9: ...igure 2 6 PHM 104A Pump Dose Capabilities 100 uL PHM 104A Infusion Capabilities by Dose 100 uL Syringe Figure 2 7 PHM 104A Pump Dose Capabilities 500 uL PHM 104A Infusion Capabilities by Dose 500 uL S...

Страница 10: ...h the retainer in the pusher block and align the syringe barrel with the syringe holder being sure that the syringe tabs are aligned with the retainer 4 Press down firmly the syringe should fit snugly...

Страница 11: ...the section of tubing in the infusion solution 4 Connect a section of tubing of the required length to the Output port on the manifold shown in Figure 2 2 Manifold End of the PHM 104A using a female...

Страница 12: ...Connecting the PHM 104A pump to a new USB port will require the reinstalling of the device driver 3 Set the FLUSH REMOTE TEST switch to REMOTE 4 Set the desired pump I D using the PUMP ID pushbuttons...

Страница 13: ...everted back to the last good value used Download Click to send Syringe Dose and Rate settings to the currently selected pump If all settings are valid the Inject and Set Default buttons will be enabl...

Страница 14: ...switches shown in Figure 2 1 Top Panel with Syringe Loaded The pump ID corresponds to the Box in MED PC 3 Set the FLUSH REMOTE TEST switch to REMOTE 4 Using the 24 VDC power supply connect the 24 VOLT...

Страница 15: ...t values P104A_INJECT Updates values and begins pump injection cycle Comments Sets the infusion parameters for the pump The maximum and minimum Rates are dependent on the Syringe Size and Dose To dete...

Страница 16: ...rogram records no data and has no end criteria P104A_Init Initializes the PHM 104A pump Must be used at the beginning of the program This command only needs to be executed once P104A_PumpSetup Sets th...

Страница 17: ...Dipper are ordered It will be necessary to change one of these LeftLight 4 RightLight 5 HouseLight 7 Pump 8 A Control Variables with Assigned Aliases as Defined Var_Alias Dose ul A 0 Default 10 ul Va...

Страница 18: ...3 Syringe Size A 2 S2 PUMP CONTROL S S 2 S1 Initialize the Pump 0 1 B P104A_Init MG SHOW 6 Init Resp B S2 S2 Wait for a Response Send the Currently selected values to the Pump R LeftLever C P104A_Pum...

Страница 19: ...efault parameters using the pump control application described in Figure 3 4 PHM 104A Control Software Screen 4 Connect the MED OPERATE port Figure 2 3 on the pump to any available output on a standar...

Страница 20: ...ation again if the PHM 104A is connected to a different USB port at a later time Insert the PHM 104A CD into the CD ROM drive and the screen shown in Figure 4 1 PHM 104A Main Screen will appear Click...

Страница 21: ...ctor on the PHM 104A to a standard wall outlet Using the included USB cable connect the USB port on the PHM 104A to any available USB port on the computer The Found New Hardware Wizard screen will app...

Страница 22: ...N Please contact Med Associates Inc for information regarding any of our products For Technical questions email support med associates com For Sales questions email sales med associates com Visit our...

Отзывы: