11.0  Technical specifications

This apparatus complies with Directive 93/42/EEC:

  Class I

Class according to EN 60601-1:

 

II

 

Type B

 

IP 20 (Recharging unit)

 

IP 20 (Silverlight)

Power supply for the recharging unit: 

 

92-255 VAC 50/60 Hz  15 VA.

Power supply for the silverlight handpiece: 

 

Rechargeable Lithium-ion battery, 3.7 V

Handpiece for intermittent operation:

 

120” ON  40” OFF.

Source of light: 

 

High-luminosity LED with optics.

 

Light radiation: 440 - 480 nm.

 

Average life: 1,800,000 cycles of 20 seconds each.

Fiber optic included in the supply: 

Diameter 8 mm.

 

Composition: Drawn coherent fibers melted in 
transparent quartz.

 

Sterilizable by steam autoclave. 

 

Minimum sterilization parameters:

 

Three times pre-vacuum cycle at 132°C for 4 
minutes. Drying time. 10 minutes.

 

(max. temp. 135°C – max. 500 cycles)

Exposure:

 

FAST:

 Exposure time 10 seconds

 

Acoustic signals indicating start and end of 

 

exposure.

 

SLOW RISE:

 Exposure time 20 seconds

 

Acoustic signal at the start, after 10 seconds and at 

the end of the 20 seconds.

 

The cycles can be stopped or repeated at any time

Time required to recharge a dead battery:

 

About two hours.

Operating conditions:

 

From +10°C to +40°C

 

Relative humidity from 30% to 75%

Transport and storage conditions:

 

From -10°C to +70°C

 

Relative humidity from 10% to 90%

 

Air pressure P: 500 hPa/1060 hPa

Weights and dimensions:

 

Recharging unit: Weight 555 g.

 

 

96 x 120 x 58 mm.

 

Silverlight handpiece: Weight 105 g.

 

 

L 198 mm  Max. Ø 22.5 mm.

Содержание Silverlight

Страница 1: ...Manual of use and maintenance...

Страница 2: ...l 10 05 4 Battery failed signal 10 05 5 Battery problems 10 05 6 Safety requirements relating to the battery 10 06 0 Use 11 06 1 Connecting the accessories 11 06 2 Safety requirements during use 12 06...

Страница 3: ...dated to the date of publication indicated on the last page The manufacturer Mectron S p A is committed to continuous updating of products which may entail changes to components of the equipment If th...

Страница 4: ...Do not use this equipment on patients fitted with pace makers or any other implantable electronic devices This requirement applies equally to the operator DANGER Point the beam of light directly at t...

Страница 5: ...R Do not install the equipment anywhere there is a risk of explosions The equipment cannot function in places where there is an inflammable atmosphere anesthetic mixtures oxygen etc DANGER Do not use...

Страница 6: ...re ensures proper functioning and reliability of all components 03 0 Delivery 03 1 Delivery of the device The equipment contains electronic components that may be damaged by impacts even inside the pa...

Страница 7: ...verlight recharging unit Fig 3 Ref A B Silverlight handpiece with rechargeable lithium ion battery Fig 3 Ref B C Fiber optic Fig 3 Ref C D Optical protection Fig 3 Ref D E Power supply cable for the r...

Страница 8: ...GER Do not short circuit the electric contacts of the recharging unit with metal objects Fig 5 Ref B and do not touch them with your hands while the apparatus is switched on WARNING The apparatus is t...

Страница 9: ...aced on the recharging unit the green battery LED does not switch on This is normal Description of the acoustic signals of the handpiece Table 2 Function Push button control Acoustic signal FAST polym...

Страница 10: ...eck the battery continuously and maintain the optimum battery charging parameters The handpiece may therefore be placed back into the recharging unit at the end of each treatment and left there regard...

Страница 11: ...wer LED will light up 05 5 Battery problems To replace a failed battery contact GC America Inc Customer Service Do not attempt to open remove change replace or in any other way adjust the battery Thes...

Страница 12: ...circuit the battery terminals This could cause burns and fire 06 0 Use 06 1 Connecting the accessories DANGER Check the condition of the device before the treatment Before each treatment always make s...

Страница 13: ...t efficient this will reduce the intensity of the light being emitted considerably In such cases it should therefore be replaced 06 3 Instructions for use The silverlight can be used in two different...

Страница 14: ...the working luminous flux measured Green luminous flux suitable for effective treatment Red luminous flux insufficient WARNING If the working luminous flux is not sufficient do not proceed with the t...

Страница 15: ...fection disconnect the recharging unit from the main power supply DANGER The casing of the recharging unit is not protected against the entry of liquids DANGER Do not spray liquids directly onto its s...

Страница 16: ...ilization procedures dry heat radiation ethylene oxide gas low temperature plasma etc DANGER The handpiece and recharging unit must not be sterilized DANGER Infection control Sterilizable parts To avo...

Страница 17: ...he battery must be disposed of and treated as waste requiring separate collection This equipment must be disposed of and treated as waste requiring separate collection Dispose of unserviceable batteri...

Страница 18: ...is ON Battery failed Contact GC America Inc Customer Service See point 05 4 and 05 5 There is no beam of light when the push button of the Silverlight is pressed and an acoustic signal is heard two be...

Страница 19: ...ed in transparent quartz Sterilizable by steam autoclave Minimum sterilization parameters Three times pre vacuum cycle at 132 C for 4 minutes Drying time 10 minutes max temp 135 C max 500 cycles Expos...

Страница 20: ...environment Guidance RF emissions CISPR 11 Group 1 The Silverlight uses RF energy only for its internal function Therefore its RF emissions are very low and are not likely to cause any interference i...

Страница 21: ...ins power qual ity should be that of a typical commercial or hospital environment Surge IEC 61000 4 5 1 kV differential mode 2 kV common mode 1 kV differential mode 2 kV common mode Mains power qual i...

Страница 22: ...gths from fixed RF transmit ters as determined by an electromagnetic site survey a should be less than the com pliance level in each frequency rangeb Interference may occur in the vicinity of equipmen...

Страница 23: ...er of transmitter W Separation distance according to frequency of transmitter m 150 kHz to 80 MHz d 1 2 P 80 MHz to 800 MHz d 1 2 P 800 MHz to 2 5 GHz d 2 3 P 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1...

Страница 24: ...her than GC America Inc authorized personnel The warranty is valid only if GC America Inc is notified within thirty 30 days following discovery of a defect For returning procedure make reference to th...

Страница 25: ...the equipment in its original packaging to ensure it is not damaged during shipment DANGER All the products must be cleaned and sterilized before returning GC America Inc will not accept and process p...

Страница 26: ...063 Fax 800 GCFAXME Caution US Federal Law restricts this device to sale by or on the order of a dentist silverlight V USA Rev 01del 02 18 2010 manufacturer Mectron S p A Via Loreto 15 A 16042 Carasco...

Отзывы: