PROBLEM
POSSIBLE CAUSE
SOLUTION
The recharging unit does not
switch on (none of the LEDs will
light up).
The power cable is not connected
correctly.
The power cable is faulty.
The recharging unit is out of
order.
Connect the cable both to the
recharging unit and to the wall
socket.
Replace the power cable.
Contact GC America Inc.
Customer Service
The red battery LED (check) of
the recharging unit is ON.
The contacts of the
recharging unit have been short-
circuited.
See point “06.6”.
The red battery LED (check) of
the recharging unit is ON.
Battery failed.
Contact GC America Inc.
Customer Service.
See point “05.4” and “05.5”.
There is no beam of light when
the push button of the Silverlight
is pressed and an acoustic signal
is heard (two beeps).
Battery dead.
Recharge the battery. See point
“05.3”.
An acoustic signal is heard at the
end of the exposure cycle
(two beeps).
Battery low.
Recharge the battery. See point
“05.2”.
An acoustic signal (three beeps)
is heard during the exposure
cycle and at the end of the cycle
Silverlight will not enable any fur
-
ther treatment to be carried out.
The thermal protection has been
activated.
It will be possible to use the
device only after it has cooled
down.
The polymerisation is insufficient.
The surface of the tip of the fiber
optic is soiled.
See point “07.4”.
The green battery LED does not
light up when the handpiece is
placed back on the charging unit.
Recharging complete. Battery
charged.
See point 04.2 - Table 1.