06.5  Safety protection

In the event of extreme heady duty use, with long and repeated exposure cycles, a thermal protection 
device is triggered automatically. 
An acoustic signal (three beeps) will be heard. 
This protection device will temporarily prevent use of the lamp for a few minutes.

06.6  Red battery LED on the recharging unit illuminated

The red battery LED (check) on the recharging unit indicates the following:
1.  Battery failed (see point 05.4).
2.  Contacts on the recharging unit short-circuited.

In the second case, to restore correct functioning of the charging unit:
1.  Disconnect the charging unit from the power supply. All LEDs extinguished
2.   Eliminate the cause of the short circuit
3.  Re-connect the charging unit to the power supply - green power LED on.

07.0  Cleaning, disinfection, sterilization

07.1  Cleaning and disinfection of the casing of the recharging unit

 DANGER: Switch off the recharging unit.

Before carrying out any cleaning and disinfection, disconnect the recharging unit from the main power 
supply.

 DANGER: The casing of the recharging unit is not protected against the entry of liquids.

 DANGER: Do not spray liquids directly onto its surface or onto the electrical contacts of the 

recharging unit.

 DANGER: The recharging unit must not be sterilized.

Proceed as follows after each treatment:

Remove the handpiece from the recharging unit.

Clean and disinfect the surface of the casing using a cloth moistened with a solution of mild 

 

detergent/disinfectant having a neutral pH (pH = 7). 

 

Follow carefully the instructions provided by the manufacturer of the disinfectant solution. 

 

Allow the disinfectant solution to air dry before reconnecting the recharging unit to the power supply. 

 

Above all, make sure that the electric contacts are completely dry.

NOTE:

 Water-based disinfectants with a neutral pH are strongly recommended. Some alcohol-based 

disinfectant solutions may be harmful and cause damage to plastic materials.

07.2  Cleaning and disinfecting the Silverlight handpiece

 DANGER: The handpiece is not protected against the penetration of liquids.

 DANGER: Do not spray liquids directly onto its surface or onto the electric contacts.

Содержание Silverlight

Страница 1: ...Manual of use and maintenance...

Страница 2: ...l 10 05 4 Battery failed signal 10 05 5 Battery problems 10 05 6 Safety requirements relating to the battery 10 06 0 Use 11 06 1 Connecting the accessories 11 06 2 Safety requirements during use 12 06...

Страница 3: ...dated to the date of publication indicated on the last page The manufacturer Mectron S p A is committed to continuous updating of products which may entail changes to components of the equipment If th...

Страница 4: ...Do not use this equipment on patients fitted with pace makers or any other implantable electronic devices This requirement applies equally to the operator DANGER Point the beam of light directly at t...

Страница 5: ...R Do not install the equipment anywhere there is a risk of explosions The equipment cannot function in places where there is an inflammable atmosphere anesthetic mixtures oxygen etc DANGER Do not use...

Страница 6: ...re ensures proper functioning and reliability of all components 03 0 Delivery 03 1 Delivery of the device The equipment contains electronic components that may be damaged by impacts even inside the pa...

Страница 7: ...verlight recharging unit Fig 3 Ref A B Silverlight handpiece with rechargeable lithium ion battery Fig 3 Ref B C Fiber optic Fig 3 Ref C D Optical protection Fig 3 Ref D E Power supply cable for the r...

Страница 8: ...GER Do not short circuit the electric contacts of the recharging unit with metal objects Fig 5 Ref B and do not touch them with your hands while the apparatus is switched on WARNING The apparatus is t...

Страница 9: ...aced on the recharging unit the green battery LED does not switch on This is normal Description of the acoustic signals of the handpiece Table 2 Function Push button control Acoustic signal FAST polym...

Страница 10: ...eck the battery continuously and maintain the optimum battery charging parameters The handpiece may therefore be placed back into the recharging unit at the end of each treatment and left there regard...

Страница 11: ...wer LED will light up 05 5 Battery problems To replace a failed battery contact GC America Inc Customer Service Do not attempt to open remove change replace or in any other way adjust the battery Thes...

Страница 12: ...circuit the battery terminals This could cause burns and fire 06 0 Use 06 1 Connecting the accessories DANGER Check the condition of the device before the treatment Before each treatment always make s...

Страница 13: ...t efficient this will reduce the intensity of the light being emitted considerably In such cases it should therefore be replaced 06 3 Instructions for use The silverlight can be used in two different...

Страница 14: ...the working luminous flux measured Green luminous flux suitable for effective treatment Red luminous flux insufficient WARNING If the working luminous flux is not sufficient do not proceed with the t...

Страница 15: ...fection disconnect the recharging unit from the main power supply DANGER The casing of the recharging unit is not protected against the entry of liquids DANGER Do not spray liquids directly onto its s...

Страница 16: ...ilization procedures dry heat radiation ethylene oxide gas low temperature plasma etc DANGER The handpiece and recharging unit must not be sterilized DANGER Infection control Sterilizable parts To avo...

Страница 17: ...he battery must be disposed of and treated as waste requiring separate collection This equipment must be disposed of and treated as waste requiring separate collection Dispose of unserviceable batteri...

Страница 18: ...is ON Battery failed Contact GC America Inc Customer Service See point 05 4 and 05 5 There is no beam of light when the push button of the Silverlight is pressed and an acoustic signal is heard two be...

Страница 19: ...ed in transparent quartz Sterilizable by steam autoclave Minimum sterilization parameters Three times pre vacuum cycle at 132 C for 4 minutes Drying time 10 minutes max temp 135 C max 500 cycles Expos...

Страница 20: ...environment Guidance RF emissions CISPR 11 Group 1 The Silverlight uses RF energy only for its internal function Therefore its RF emissions are very low and are not likely to cause any interference i...

Страница 21: ...ins power qual ity should be that of a typical commercial or hospital environment Surge IEC 61000 4 5 1 kV differential mode 2 kV common mode 1 kV differential mode 2 kV common mode Mains power qual i...

Страница 22: ...gths from fixed RF transmit ters as determined by an electromagnetic site survey a should be less than the com pliance level in each frequency rangeb Interference may occur in the vicinity of equipmen...

Страница 23: ...er of transmitter W Separation distance according to frequency of transmitter m 150 kHz to 80 MHz d 1 2 P 80 MHz to 800 MHz d 1 2 P 800 MHz to 2 5 GHz d 2 3 P 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1...

Страница 24: ...her than GC America Inc authorized personnel The warranty is valid only if GC America Inc is notified within thirty 30 days following discovery of a defect For returning procedure make reference to th...

Страница 25: ...the equipment in its original packaging to ensure it is not damaged during shipment DANGER All the products must be cleaned and sterilized before returning GC America Inc will not accept and process p...

Страница 26: ...063 Fax 800 GCFAXME Caution US Federal Law restricts this device to sale by or on the order of a dentist silverlight V USA Rev 01del 02 18 2010 manufacturer Mectron S p A Via Loreto 15 A 16042 Carasco...

Отзывы: