page 9
Press the
MAX
key again, and the word
MAX
has now
disappeared from the display. The display will now indicate
torque of either direction, as it is applied, maintaining a
'running' display.
Press the
RESET
key to clear both maximum registers and
prepare for detecting the next maximum readings.
The Orbis has RS232, Digimatic and analogue output
signals. It is possible to transmit the displayed reading to
peripheral devices (e.g. PC, printer) by pressing and
releasing the
TXD
key.
Displayed readings can also be requested individually from
a PC via the RS232 interface by sending a "?" character. See
page 13. When using the Orbis for creating a graphical
presentation of data, ensure that the normal mode, and not
a max mode, is selected.
For sending a continuous data stream to a PC, press and
hold the
TXD
key for 2 seconds. TX will now appear in the
display to indicate that data is being sent. To stop sending
data, simply press and release the
TXD
key, at which point
TX will disappear from the display.
Please note that the continuous data stream only starts
when approximately 2% of the rated capacity of the Orbis
is reached.
'Normal' mode
Data Output
Orbis uses 9600 Baud 8 data
bits, 1 start bit 1 stop bit and
no parity.
A full range of data cables are
available to connect the Orbis
to peripheral devices. Please
contact your supplier for
further information.