page 6
*
Do not overload
the torque sensor, as this could cause
irreparable damage.
Loads greater than 120% of full-scale will produce an
audible beep until load is released and an OL symbol will
appear on the display
for 30 seconds
.
Loads greater than 150% of full-scale will produce an
audible beep until load is released and an OL symbol will
appear
permanently
on the display. Consult your supplier
to arrange repair.
To power down the Orbis press the red
key.
Clockwise torque is displayed on the Orbis and recognised
by the symbol shown in Fig. 2.
Counter-clockwise torque is displayed on the Orbis and
recognised by the symbol shown in Fig. 3c.
A torque indicator bar alerts the operator to how much
load has been applied to the torque sensor. As the load
approaches the maximum rating of the torque sensor, the
indicator bar changes appearance when above
approximately 80% of the rated capacity. This warns the
operator that steps should be taken to prevent excessive
torque being applied.
Fig. 2
Clockwise and
Counter-clockwise Values
If the Orbis has suffered a
serious overload condition, the
torque indicator bar will be
partially displayed even when
no torque is present. This is a
warning that the torque sensor
is damaged and you should
immediately contact your
supplier to arrange repair.
Symbol for
clockwise
torque
Unit of
measurement
Load indicator