8
Mecmesin
Assembly and Installation of –
i
System Test Stands
3.6.1
Limits after connection to the PC, with the software running
If the crosshead reaches a limit during a test, it will stop then reverse direction for
approximately 5 mm, then stop again. When this happens an onscreen warning message
will be displayed:
Click
OK
to acknowledge the warning and use the Jog buttons to move the crosshead so
that you can then check the position of the limit stops and re-set if required.
If a limit stop is reached while pressing one of the Jog buttons on the front panel of the
MultiTest-
i
a different warning message is displayed.
Clicking
OK
will allow you to correct the situation and continue.
Clicking
Abort
will disconnect the stand and display the following message:
Click
OK
then
Exit
to go to the Front Screen, this will reconnect the stand again.