USB-5106 User's Guide 

Functional Details 

13 

Voltage input adapter cables 

A voltage input adapter cable is a resistive voltage divider that allows a voltage input to be recorded. The  
USB-5106 supports the following voltage input adapter cables: 

 

CABLE-ADAP5 measures 0 V to 5 VDC 

 

CABLE-ADAP10 measures 0 V to 10 VDC 

 

CABLE-ADAP24 measures 0 V to 24 VDC  

Voltage input cables are for use in indoor environments only. 

Figure 7 shows a schematic of the voltage input adapter cable. All resistors are 0.1%. 

 

Figure 7. Voltage input adapter cable diagram 

Reading out the logger 

To read out data logged by the USB-5106, connect it to the computer with the USB cable and read out it with 
USB-5100 Series Data Logger Software. 

Refer to the 

USB-5100 Series Software Help

 for detailed information about how to read out logged data. 

Recording internal logger events 

The logger records the following internal events to track logger operation and status. You can plot these events 
after reading out the logger and opening the data file. 

Internal Event Name 

Definition 

Host Connected 

The logger is connected to the computer. 

Started 

The Start/Stop button was pressed to begin logging. 

Stopped 

The logger received a command to stop recording data (from either the USB-5100 
Series Data Logger Software or by pushing the 

Start/Stop

 button). 

Button Up/Button Down 

The 

Start/Stop

 button was pressed for 1 second. 

Chan <#> Alarm Tripped 

An alarm has tripped on the specified channel. 

Chan <#> Alarm Cleared 

An alarm has cleared on the specified channel. This event also contains the value that 
is furthest out of range for the sensor before the alarm cleared. 

New Interval 

The logger has entered or exited burst logging mode. 

Safe Shutdown 

The battery level dropped below 1.85 V; the logger performs a safe shutdown. 

Содержание USB-5106

Страница 1: ...USB 5106 4 channel analog logger User s Guide Document Revision 2A February 2015 Copyright 2015 ...

Страница 2: ...in any form by any means electronic mechanical by photocopying recording or otherwise without the prior written permission of Measurement Computing Corporation Notice Measurement Computing Corporation does not authorize any Measurement Computing Corporation product for use in life support systems and or devices without prior written consent from Measurement Computing Corporation Life support devic...

Страница 3: ...nctional Details 7 External components 7 Start Stop button 7 Alarm Stats button 7 Mounting magnets 7 Battery compartment 7 USB port 7 Mounting loops 7 External sensor inputs 8 LCD Screen 8 Setting up the USB 5106 9 Data logging options 9 Logging modes 9 Alarm modes 10 Connecting external sensors 11 Compatible sensors 11 Reading out the logger 13 Recording internal logger events 13 Mounting the log...

Страница 4: ...of objects on a screen such as buttons text boxes and check boxes italic text Italic text is used for the names of manuals and help topic titles and to emphasize a word or phrase Where to find more information Additional information about USB 5106 hardware is available on our website at www mccdaq com You can also contact Measurement Computing Corporation with specific questions Knowledgebase kb m...

Страница 5: ...ip when the sensor reading rises above or falls below a specified measurement Set up burst logging in which the logger records data at a different interval during certain conditions Calculate minimum maximum average and standard deviation statistics Use the built in LCD screen to check current readings and to monitor logging status battery use and memory consumption between readouts TheUSB 5106 da...

Страница 6: ...device is installed when you install the software Therefore you need to install the USB 5100 Series Data Logger Software before you install the hardware For operation on a Windows operating system we recommend that you run Windows Update to update your operating system with the latest USB drivers Perform the following steps to connect a USB 5100 Series data logger 1 Install the two AAA batteries S...

Страница 7: ...er events on page 13 or to turn on the LCD screen on if it is configured to be off Alarm Stats button Use this button to clear a tripped alarm see Alarm on page 10 or to cycle between statistics alarm readings and the current sensor reading Mounting magnets Four built in mounting magnets on the rear of the device are used to mount the logger to a magnetic surface Battery compartment The battery co...

Страница 8: ...below The alarm symbol clears depending on how alarms are configured If the alarm is configured to clear when the logger is relaunched this symbol remains on the LCD Otherwise it clears when the sensor reading is back within the alarm limits or you press the Alarm Stats button for 3 seconds The channel number associated with the sensor reading Up to four channels are visible at one time The sensor...

Страница 9: ...elected in the software If you choose a logging interval less than 15 seconds the data will be recorded at the faster interval but the temperature readings will only be updated on the screen every 15 seconds When the logger has stopped logging the LCD screen remains on until the logger is offloaded to a computer unless launched with the Turn LCD off option Once the logger is offloaded and disconne...

Страница 10: ...nsor channels The alarm symbol displays on the LCD screen when an alarm is tripped Alarm limits are checked when the LCD screen refreshes every 15 seconds You can set the alarm mode to Cumulative or Consecutive Cumulative raises an alarm when the sensor is out of range for a specified number of samples The high or low samples don t need to be consecutive Consecutive raises an alarm when the sensor...

Страница 11: ...rnal input jacks but others require additional cables as described in the following sections If you disconnect a sensor or if it is not fully inserted into the jack while the logger is logging an erroneous sensor reading can appear on the LCD for that channel In addition erroneous readings will be logged and saved in the data file depending on the logging interval for example if a sensor is discon...

Страница 12: ...ut off the end of the gray cable where it connects to the blue and yellow wires as it contains the precision resistor required for current measurement The 4 20 mA input cable is for use in indoor and outdoor environments Figure 5 shows a schematic of the 4 20 mA input cable Figure 5 4 20 mA input cable diagram Figure 6 shows a wiring diagram of the 4 20 mA input cable Figure 6 4 20 mA input cable ...

Страница 13: ...how to read out logged data Recording internal logger events The logger records the following internal events to track logger operation and status You can plot these events after reading out the logger and opening the data file Internal Event Name Definition Host Connected The logger is connected to the computer Started The Start Stop button was pressed to begin logging Stopped The logger received...

Страница 14: ...kV Avoid electrostatic discharge by grounding yourself using a wrist strap or touch the computer chassis or other grounded object to eliminate any stored static charge Battery information The logger requires two user replaceable AAA 1 5 volt alkaline batteries Non rechargeable lithium batteries can be used for operation at the extreme ends of the logger operating range Expected battery life varies...

Страница 15: ...s in fire Do not expose the contents of the batteries to water Dispose of the batteries according to local regulations The USB 5100 Series Data Logger Software lets you record the current battery voltage at each logging interval this option is disabled by default Recording battery life at each logging interval takes up memory and therefore reduces logging duration It is recommended you only record...

Страница 16: ...hen memory full push button or date time Restart mode Push button Time accuracy 1 minute per month at 25 C 77 F refer to the Time Accuracy plot below Resolution 16 bit Battery life 1 year typical with logging rate of 1 minute and a sampling interval of 15 seconds or greater Battery type Two AAA 1 5 V alkaline batteries user replaceable Memory 4 MB 1 9 million measurements maximum Download type USB...

Страница 17: ...le specifications Parameter CABLE 4 20mA CABLE ADAP5 CABLE ADAP10 CABLE ADAP24 Measurement Range 0 mA to 20 1 mA 0 V to 5 0 V 0 V to 10 V 0 V to 24 V Accuracy 0 001 mA 0 2 of reading 0 2 mV 0 3 of reading 0 4 mV 0 3 of reading 1 0 mV 0 3 of reading Resolution 0 3 µA 80 µV 160 µV 384 µV ...

Страница 18: ...Measurement Computing Corporation 10 Commerce Way Suite 1008 Norton Massachusetts 02766 508 946 5100 Fax 508 946 9500 E mail info mccdaq com www mccdaq com ...

Отзывы: