PCI-DAS6052 User's Guide
Specifications
Accuracy
333 kS/s sampling rate, single channel operation and a 15-minute warm-up. Accuracies
listed are for measurements made following an internal calibration. They are valid for
operational temperatures within ±1°C of internal calibration temperature and ±10°C of
factory calibration temperature.
Calibrator test source high side tied to Channel 0 High
and low side tied to Channel 0 Low. Low-level ground is tied to Channel 0 Low at the
user connector.
Table 6-1. Absolute Accuracy
Range Absolute
Accuracy
±10V
±15.6 LSB
±5V
±5.7 LSB
±2.5V ±15.6
LSB
±1V ±15.7
LSB
±500mV
±15.9 LSB
±250mV ±18.0
LSB
±100mV ±21.0
LSB
±50mV
±23.0 LSB
0 to 10V
±8.1 LSB
0 to 5V
±27.8 LSB
0 to 2V
±28.0 LSB
0 to 1V
±28.0 LSB
0 to 500mV
±31.7 LSB
0 to 200mV
±36.4 LSB
0 to 100mV
±38.7 LSB
6-2