Chapter 5
94
Testing Single-Value Analog Input
To verify that the board can read a single analog input value, do the following:
1.
Connect a voltage source, such as a function generator, to analog input channel 0
(differential mode) on the DT3010 Series board. Refer to
for an example of how to
connect a differential analog input.
2.
In the Quick DataAcq application, choose
Single Analog Input
from the
Acquisition
menu.
3.
Select the appropriate DT3010 Series board from the Board list box.
4.
In the Channel list box, select analog input channel 0.
5.
In the Range list box, select the range for the channel.
The default is ±10 V.
6.
Select
Differential
.
7.
Click
Get
to acquire a single value from analog input channel 0.
The application displays the value on the screen in both text and graphical form.
Содержание Data Translation DT3010 Series
Страница 2: ...DT3010 Series UM 16866 V User s Manual Title Page ...
Страница 5: ......
Страница 15: ...About this Manual 14 ...
Страница 16: ...15 1 Overview Features 16 Supported Software 18 Accessories 19 Getting Started Procedure 21 ...
Страница 23: ...Chapter 1 22 ...
Страница 24: ...Part 1 Getting Started ...
Страница 25: ......
Страница 37: ...Chapter 2 36 ...
Страница 49: ...Chapter 3 48 ...
Страница 91: ...Chapter 4 90 ...
Страница 102: ...Part 2 Using Your Board ...
Страница 103: ......
Страница 147: ...Chapter 6 146 ...
Страница 159: ...Chapter 7 158 ...
Страница 160: ...159 8 Calibration Calibrating the Analog Input Subsystem 161 Calibrating the Analog Output Subsystem 169 ...
Страница 176: ...175 9 Troubleshooting General Checklist 176 Technical Support 178 If Your Board Needs Factory Service 179 ...
Страница 181: ...Chapter 9 180 ...
Страница 193: ...Appendix A 192 ...
Страница 213: ...Appendix B 212 ...
Страница 231: ...Index 230 ...