Installation

Prior to installing the flow meter, remove the storage caps from the ports and look carefully into each port of the meter. 
Ensure that no dirt or foreign particles have gotten into the ports of the meter. Make sure that adequate filtration 
exists upstream of the flow meter and that no contaminants exist in the line between the filter and flow meter. It is 
recommended that the flow meter be connected to the circuit by means of unions close to the flow meter to allow easy 

removal.

A by-pass valve should be installed between the inlet and outlet ports in parallel to the flow meter to allow flow through 
the system in the event that the flow meter becomes blocked by foreign material.

 

The preferred orientation for the meter is to place the transmitter to the side of the flow meter.  Such an installation 

ensures that bubbles will not accumulate within the meter. In a high pressure installation, the compression and 
expansion of gas bubbles from the bearing pockets may create undesirable side forces on the gears.  Mounting the 
meter with the transmitter to the side will purge the air from the meter.  This orientation is also preferred as it reduces 
the amount of heat that rises from the meter into the transmitter circuitry.

The following items and conditions should be considered:

Line and Bypass Valves: These valves allow filter cleaning or flow meter removal without completely shutting the system 

down and draining the lines. They also allow system start up under conditions which could damage the meter; such as: 
air in the lines, high temperature materials, or initial line surges.

Filtration: Clearances between the gears and internal wall are typically 0.001” to 0.002”. Any dirt present in the system 
can jam or damage the unit. A 10 micron filter (such as a Max 381 Series stainless steel unit) is generally recommended, 
although materials with very high viscosities may require a coarser filter. For bidirectional flow applications, use a filter 
on each side of the flow meter. Materials with fibrous or non abrasive particulate matter may have to be run without 
filters. Follow the recommendation of your Max Sales Engineer or consult our Technical Service Department.

Inlet and Outlet Ports: Use the “IN” port as the inlet for the most predominant flow direction. Install the flow meter on 

the discharge side of the pump whenever possible. Excessive vibration at the meter should be avoided.

High Temperatures: Mount the meter so that the transmitter is below or to the side of the meter. This minimizes heat 

transfer by convection from the flow meter to the transmitter. The transmitter is the most heat sensitive element in the 
system and the transmitter manual should be consulted for specific limits. An optional fluid heater block can be used 
on the flow meter to keep it at operating temperature during standby conditions. For substances that are solid at room 
temperature, the block may be required to keep the material molten and flowing through the meter.

Clean Plumbing: Before installing the flow meter, clean the inside of the pipe line with compressed air or steam 

(especially when using new pipe). Don’t use water, steam, or compressed air on the meter itself!

Содержание G004

Страница 1: ...Model G004 Model G045 Model G105 Model G240 Model G015 G Series Positive Displacement Gear Flow Meters Manual G004 G015 G045 G105 G240 G004HZ G015HZ G045HZ G105HZ G240HZ...

Страница 2: ...stall page 3 Meter General Description page 4 Transmitter General Description page 5 Meter Specifications page 5 Transmitter Specifications page 5 Transmitter Specifications Temp page 6 Available Desi...

Страница 3: ...plumbing supplies Many Max meters are installed and operate for decades so having the following information in your records may prove useful We have provided this outline as a starting point Process T...

Страница 4: ...ter Available in 303 and 316 stainless steel Transmitter Features High resolution measurement Analog Output Configured output ranges to any value within 10 Vdc or 20 mA Frequency Output Configured out...

Страница 5: ...e Default 50 Revolution of Meter for unidirectional software selectable from 1 100 of 1 revolution 2 for bi directional meters Signal Filtering Software selectable from 1ms to 250ms time constant 1 Co...

Страница 6: ...ature graph Model 29X Transmitter Series 25 0 50 20 40 60 80 100 120 100 150 200 250 300 Process Temperature C Ambient Temperature C 225 110 Ultra High temperature 2 part pickup Standard High temperat...

Страница 7: ...allow system start up under conditions which could damage the meter such as air in the lines high temperature materials or initial line surges Filtration Clearances between the gears and internal wal...

Страница 8: ...stem Flowing near the meters maximum flow rate for a given viscosity will purge air bubbles Tilting tapping or shaking the meter at lower flow rates will also dislodge entrapped air Run water or aqueo...

Страница 9: ...on 1 The transmitter is attached to the flow meter s threaded magnet shield Hand tighten only 3 ft lb 2 The transmitter lid has four thread paths To realign the cable remove the lid and rotate up to 1...

Страница 10: ...th S Turck Connector PCA Label Pin Mating Cable Wire Color Power V 1 Brown Common Com 4 Black Signal Output Sig 5 Grey Signal Output Ret 2 White Case Ground Case 3 Blue 2 1 4 5 3 Voltage or Current An...

Страница 11: ...Max Machinery Inc G Series User Manual Copyright 2018 Rev2018A 11 Model G015 Flow Meter Model G004 Flow Meter Model G045 Flow Meter Pressure Drop Charts...

Страница 12: ...Max Machinery Inc G Series User Manual Copyright 2018 Rev2018A 12 Model G240 Flow Meter Model G105 Flow Meter Pressure Drop Charts continued...

Страница 13: ...furbishment will be provided upon request for quotation submitted online via https www maxmachinery com service request Max Machinery Inc 33A Healdsburg Ave Phone 707 433 2662 Healdsburg CA 95448 www...

Страница 14: ...owered by 24 Vdc 15 Liters min G015 A B 4 20mA output powered by 12 Vdc 45 Liters min G045 A C 0 10 Volt output powered by 24 Vdc 105 Liters min G105 A D 0 10 Volt output powered by 12 Vdc 240 Liters...

Страница 15: ...Liters min G004 A A 4 20mA output powered by 24 Vdc 15 Liters min G015 A B 4 20mA output powered by 12 Vdc 45 Liters min G045 A C 0 10 Volt output powered by 24 Vdc 105 Liters min G105 A D 0 10 Volt o...

Страница 16: ...rder 5 This limited warranty covers all defects encountered in normal use of Seller s products only and does not apply to the following cases a Disassembly of the meter b Failure to install recommende...

Страница 17: ...KIND AGAINST SELLER BE GREATER IN AMOUNT THAN THE PURCHASE PRICE OF THE PRODUCT SOLD BY SELLER CAUSING THE ALLEGED DAMAGE WITHOUT LIMITING THE FOREGOING PURCHASER ASSUMES ALL RISK OF LIABILITY FOR LO...

Отзывы: