7. Operation
The display window on the Flow Monitor panel functions as a measurement display window (for
checking measured values) and a setting window (for entry of parameters), and must be operated in
accordance with each situation.
The window operation procedures and indications on display are described below.
<Measurement Display Window>
<Setting Window>
Displays measurement results.
(Flow rate in % is shown.)
Displays measurement results.
(Vibration amplitude V is shown.)
Displays measurement results.
(Frequency Hz is shown.)
Displays arithmetic processing results.
(The product of
vibration amplitude V frequency Hz is shown.)
Press [STO].
Press [STO].
Press [STO].
Press [STO].
Displays measurement results.
(Flow rate in % is shown.)
Displays the current zero setting.
(Enter a value to be set for 0%.)
Input value range: 0 – 9999
Default: 0
Displays the current span setting.
(Enter a value to be set for 100%.
Input value range: 10 – 57000
Default: 57000
Displays the current moving-average setting.
(Enter a value for the
number of times the moving average is taken.)
Input value range: 1 – 90 times
Default: 5 times
Press
[STO].
Press
[STO].
Press
[STO].
Press
[STO].
Press [MOD].
Press [STO].
Press [MOD].
Press [MOD].
Press [MOD].
Displays the current sampling time setting for automatic entry
and setting operation.
(Enter a value for the sampling time.)
Input value range: 1 – 60 minutes
Default: 1 minute
Note: Press the [STO] button to return from the setting window to the measurement display window.