- 10 -
7-1-2. Automatic Entry and Setting Method
With the automatic entry and setting method, data measured within the sampling time (which
has been set to any desirable value within the setting range) is calculated. The zero and span
settings are automatically entered and stored to reflect the calculation results.
Note: The default setting for the sampling time is one minute.
(a)
Setting Zero
(b)
Setting Span
Displays measurement results.
(Flow rate in % is shown)
Displays the current zero setting.
All the numeric characters will flicker.
After the value for zero is set, the window will automatically
return to being the measurement display window.
Press [MOD].
Hold down [MOD] for
three seconds or longer.
Displays measurement results.
(Flow rate in % is shown.)
Displays the current span setting.
All the numeric characters will flicker.
After the value for span is set, the window will automatically
return to being the measurement display window.
Press [MOD] twice.
Hold down [MOD] for
three seconds or longer.