18
-
8-8.Damping Setting
【
Menu
:
1-4
】
This item specifies the time constant (processing time until the output value reaches 63.2% of
the normal output value (100%)).
See Figure 12.
When the time constant is set, the normal output value forms a curve shown in Figure 12.
If the input/output signal is fluctuated, you can increase the setting value to increase the processing
time taken to reach the normal output value, so that the fluctuation of the output signal is suppressed.
Setting range: 0 - 999 s (default value: 0 s)
Times
Signal level
100%
Time
Normal output level
63.2%
Fig.12. Time constant and primary delay element
5 0.0
%
Menu 0
▶
1. Basic setup
2. Test
3. Reset
4. Language
5. Device info
Basic setup 1
▶
1. Display
2. Relay output
3. Analog output
(S. failure mode)
4. Damping
5. Failure setup. In
5 0.0
O K ×1
O K ×1
+ ×3
ESC ×3
(Measuring screen)
O K ×1
( Use ESC button to return
to the measuring screen. )
Damping 1-4
Damping time
▶
0 0 0 s
( Setting is saved )
%
(Measuring screen)
O K ×1
Damping 1-4
Damping time
▶
0
0 0 s
*For example, the
value is
set to 30s.
Basic setup 1
1. Display
2. Relay output
3. Analog output
(S. failure mode)
▶
4. Damping
5. Failure setup. In
O K ×1
Damping 1-4
Damping time
▶
0
3
0 s
Damping 1-4
Damping time
▶
0 3 0 s
Use
▶
and + buttons to
change the value to 030s.
(
A blinking
indicates the
item to be set. )