14
-
8-4.No.1 Limit Setting
【
Menu
:
1-2-1
】
This item sets the contact output operation point ( ON / OFF point ) of the No.1 Limit by %.
4mA and 20mA input signals correspond to 0% and 100% respectively.
The set minimum hysteresis for ON and OFF points is 0.1%.
Setting range : 0.0
~
100.0
( default value : ON point
→
95.0%, OFF point
→
85.0% )
5 0.0
%
Menu 0
▶
1. Basic setup
2. Test
3. Reset
4. Language
5. Device info
Basic setup 1
▶
1. Display
2. Relay output
3. Analog output
(S. failure mode)
4. Damping
5. Failure setup. In
5 0.0
O K ×1
O K ×1
+ ×1
ESC ×4
(Measuring screen)
O K ×1
( Use ESC button to return
to the measuring screen. )
Relay output 1-2
▶
1. No.1 Limit
2. No.2 Limit
3. S. failure relay
(
A blinking
indicates the
item to be set. )
Use
▶
and + buttons to
change the value to 80.0.
O K ×1
( Setting is saved )
%
(Measuring screen)
O K ×1
No.1 Limit 1-2-1
ON point
▶
95.0 %
OFF point
85.0 %
*For example, the
ON point is changed
from "95.0" to "80.0".
O K ×1
Basic setup 1
1. Display
▶
2. Relay output
3. Analog output
(S. failure mode)
4. Damping
5. Failure setup. In
No.1 Limit 1-2-1
ON point
▶
95.0 %
OFF point
85.0 %
No.1 Limit 1-2-1
ON point
▶
8
0
.0 %
OFF point
85.0 %
No.1 Limit 1-2-1
ON point
▶
80.0 %
OFF point
85.0 %