5 Options
5-2
Matsusada Precision
AK series
5-2
Slow Start (LW)
● Select the HV ON/OFF switch to ON, and the maximum rated voltage can be attained in 10 seconds
approximately.
If the output voltage is set to 50% of the maximum rated voltage, the set voltage is attained
in 5 seconds approximately after the HV ON/OFF switch is selected to ON.
●
With the remote switch (LS) and the remote reset (LT) selected to ON, the maximum rated voltage
can be attained in 10 seconds approximately.
5-3
Overload Trip (LOc)
When the output current exceeds value set by the CURRENT dial or external voltage, the output is cut
off. During cutting off, the ”CC” LED is ON and the OUTPUT LED is OFF.
Once select the OUTPUT ON/OFF switch to OFF and next select to ON again to reset.
The external switch can reset the cut-off condition at output over current.
(The open collector can also do.)
Switch 4 Short
=
HV reset
Open
Short
Reset
5-4
AC Input 400V L(400V)
Input voltage: AC400V±10
%
, 50/60Hz, Three phase
Caution
“L” = HV reset
Fig. 5-3
Remote reset
Switch 4
TB1
16
17
18
19
20
21
3
4
5
6
7
8
9