Note: When output in OFF mode and
button is dark, rotary knob and up/down
keys can’t be used to adjust voltage and current.
If rotary knob is enabled, then adjusting it will real-time change the current output
Enter
setup without pressing
to confirm.
4.5.5 Current Operation
I-set
Solution 1
: press
(Local) to select channel, press
then enter a numerical value
Enter
followed by
.
I-set
Solution 2
: Press
, then press
to move the cursor position and adjust the
current value using the knob.Press
Esc
or
Enter
to exit.
I-set
Solution 3
: Press
, then press
to move the cursor position and adjust the
to exit.
. Press or
Esc
Enter
current value using
4.5.6 Save and Recall Operation
You can store up to 40 different operating states in memory locations 1 through 40.They are
divided to four groups, each group includes nine different setups. Each operating state
includes a constant voltage value, constant current value, maximum output voltage.
Press
(Shift)
+
Recall
(Save) followed by a number key to save the current operating
state to non volatile memory.
Press
Recall
+number 1 to 9 to recall operating state assigned to this location.
You can also use the SCPI command(*SAV
、
*RCL) to save and recall.
When Save operation is done,there will be a corresponding information to indicate the
successful or failed operation.While for Recall operation,only information to indicate failed
recall.
Note: The power supply doesn’t support Save/Recall operation when in serial/parallel
or tracking mode.
If in serial/parallel mode and do Save/Recall operation, the VFD will display INV
OPER (invalid Operation) to remind the user of a wrong operation.
If parameters want to be recalled do not exist, the current value position on VFD will
display -----, and then resume after 2S.
4.5.7 OVP operation
Select the channel-----press
(Shift)+
(OVP)-----select “ON” and continue to set
USER MANUAL
25