11
The power returned commands are "0,'No Error'", or "50,'Error Para Count'", or
"70,'Invalid Command'" and so on.
Order [:]SYSTem:REM
The command is used to set up the power supply for remote control mode, that is, PC
control mode.
Return command: None
Example: send the command to [:]
SYST:REM
<LF>
After the transmission is successful, the front panel of the power supply is set to the remote
control mode. At this time, the key on the front panel is invalid. Only by pressing the Shift+
number 7 key, or sending the command []SYST:LOC: can the power supply be switched to
the front panel of the power supply.
Order [:]SYSTem:LOC
The command is used to set the power supply to the panel control mode.
Return command: None
Example: if the power is in remote control mode (PC) at this time, send the command
[:]
SYST:LOC
<LF>
After the transmission is successful, the front panel of the power supply has been set to the
front panel control mode.
6.2.3 Measurement command
The commands in this section are all commands under [MEASure root path. The commands
given are all in a short mnemonic format.
Order []MEASure:VOLTage:?
The command is used to read the output voltage value of the power supply.
Return command: numeric value
Return parameter unit: v
Example: the order sent is: [:]
MEAS:VOLT?
<LF>
At this time, the power return command is: 12.560, which means the current value of
the power supply is 12.560v.
Order []MEASure:CURRent:?
The command is used to read the output current value of the power supply.
Return command: numeric value
Return parameter unit: a
Example: the order sent is: [:]
MEAS:CURR?
<LF>