Section 3  Controls and Indicators 

 

Pictures of the display and controls are shown at the rear of this manual. 

3.1 

Bargraph and exponent 

The bargraph and exponent indicate the leak 
rate in terms of the selected units. For 
example, a leak rate of +6 x10-4 ml/sec would 
appear as shown opposite. 

3.2 Zero 

Operating this button sets the detected value, the display, and audible signal to zero. 

3.3 Units 

Press 

ZERO

 and 

GAS GROUP

 together to step 

through the available units of measure to set the one 
desired. The correspondence between leak rate 
settings depends on the sample flow rate. If accurate measurements are required the 
instrument should be calibrated in the same mode as it will be used. 

3.4 Peak 

hold

 

 

Press 

PEAK HOLD

 to select and display the Peak Hold symbol on the screen. In this 

mode, the instrument shows the highest positive, or negative, value measured since the 
mode was entered.  Press 

PEAK HOLD

 again to exit this mode.  The audible leak 

indication remains unaffected. 

3.5  Gas group button

 

Use this button to see and select the required gas group.  
Upon pressing the button, the display will blank except for the 
Leak Rate Exponent which shows the current gas group.  
After two seconds, the displayed number will start to cycle 
through the gas groups 1 to 4. 

To keep the current gas group, press the button again immediately.  To select a 
different gas group, press the button again when the required group number is visible. 

The gas groups, and thermal conductivities, are listed in Section 8. 

INT-0124-XX                     7

 

Содержание LeakHunter Plus 8066

Страница 1: ...LeakHunter Plus 8066 User Manual LeakHunter Plus LTR REVISION DATE BY A1 Release for Prod 11 02 95 CJC B1 Revise Spares List 05 23 96 BT INT 0124 XX...

Страница 2: ...anties shall not apply if the equipment has been repaired other than by Matheson or a service facility designated by Matheson or if this equipment has not been operated and maintained in accordance wi...

Страница 3: ...conductivity that differs from that of the ambient air on which it was zeroed The instrument is normally zeroed in ambient air comprising a mixture of approximately 78 nitrogen and 21 oxygen Hence th...

Страница 4: ...nge extension nozzle Probe clip and template AC power adapter and battery charger Battery User manual Cleaning wipe Extended probe The clip attaches permanently to the rear of the instrument The probe...

Страница 5: ...n in Figure 1 3 Peel off the paper that protects the adhesive on the back of the clip Assembly aid Probe clip Rear of case Figure 2 Probe clip template 4 Carefully position the clip in the position de...

Страница 6: ...ctable Functions Peak hold Zero Calibrate Gas type nozzle type Response time Approximately 1s Recovery time Approximately 1s Audio Signal proportional to leak rate Diagnostics Battery low indication D...

Страница 7: ...d in the same mode as it will be used 3 4 Peak hold Press PEAK HOLD to select and display the Peak Hold symbol on the screen In this mode the instrument shows the highest positive or negative value me...

Страница 8: ...zzle Present symbol The same operations should be carried out when attaching or removing the extended probe supplied as part of the 8066 3 8 Battery low This symbol becomes visible when the battery is...

Страница 9: ...and hold ZERO then press PEAK HOLD You will see the displayed exponent change with the change of sensitivity For instance with ml sec units successive exponents will be 5 most sensitive range 4 and 3...

Страница 10: ...are disabled 4 Present a standard leak or concentration of gas to the instrument 5 Set the display bargraph and exponent to read the standard leak flow rate or concentration use PEAK HOLD to increase...

Страница 11: ...pter is for indoor use only The battery charger is for indoor use with Ni Cad batteries only Do not attempt to recharge other types of battery as damage or injury may result Recharging time is 16 hour...

Страница 12: ...nd probe Instrument will not power up Depleted battery Replace battery or use AC power adapter Wrong connector used on AC power adapter Make sure smaller female connector is used on unit 7 2 Error ind...

Страница 13: ...05 6 0E 04 2 1E 08 1 6E 02 2 Neon 41 5 80E 05 3 5E 03 1 2E 07 9 3E 02 Xenon 41 5 81E 05 3 5E 03 1 2E 07 9 3E 02 3 R11 35 6 84E 05 4 1E 03 1 5E 07 1 0E 03 R12 30 7 90E 05 4 7E 03 1 7E 07 1 3E 03 R21 3...

Страница 14: ...Units Peak hold Battery low Leak rate bargraph mantissa Leak rate exponent Not used Default calibration in use Nozzle present LeakHunter Plus 8066 ZERO PEAK HOLD GAS GROUP 1 2 POWER INT 0124 XX 14...

Отзывы: